Flöge J, Stannat S, Bahlmann J, Kühn K W, Koch K M
Blood Purif. 1986;4(1-3):112-9. doi: 10.1159/000169434.
In order to obtain performance data and biocompatibility characteristics of the two new Gambrane FD 100 polycarbonate hollow-fiber dialyzers (y-sterilized, dry or prefilled with isotonic saline) a paired in vivo evaluation was carried out. The measured data indicate that residual blood volume and small molecular weight clearances of Gambrane hollow-fiber dialyzers are not significantly different from comparable Cuprophan dialyzers. Small molecular clearance values are not affected by the significantly different ultrafiltration coefficients of the two tested FD 100 dialyzers. Biocompatibility of Gambrane as reflected by white blood cell count, platelet count and C3d is superior to Cuprophan.
为了获取两款新型Gambrane FD 100聚碳酸酯中空纤维透析器(经γ射线灭菌、干燥或预充等渗盐水)的性能数据和生物相容性特征,进行了一项配对体内评估。测量数据表明,Gambrane中空纤维透析器的残余血量和小分子清除率与可比的铜仿膜透析器无显著差异。小分子清除率不受两款受试FD 100透析器超滤系数显著差异的影响。由白细胞计数、血小板计数和C3d反映的Gambrane生物相容性优于铜仿膜。