In vivo analysis of the Lundia 5 (PRO 5) and Lundia 3 (PRO 3) dialyzers with Gambrane polycarbonate membranes.

作者信息

Miller J H, Shinaberger J H, Henderson L W, Gardner P W

出版信息

Blood Purif. 1986;4(1-3):63-73. doi: 10.1159/000169428.

Abstract

The physical, solute transfer and biocompatibility properties of two new parallel plate dialyzers using Gambrane polycarbonate (PC) membrane of 0.8 m2 (L-3 PC) and 1.1 m2 (L-5 PC) were studied in 6 volunteer chronic dialysis patients. Two of these 6 had suffered severe first-use syndromes with Cuprophan. Samples were collected under carefully controlled conditions. Mass transport properties were confirmed by total dialysate collection and found to compare favorably with other standard dialyzers. For the L-3 PC, KUF was 3.59 ml/h/mm Hg TMP-14.5 and standard whole blood clearances were BUN 142, creatinine 121 and inorganic phosphate 69 ml/min. For the L-5 PC, KUF was 6.27 ml/h/mm Hg TMP-28.66 and standard clearances were BUN 155, creatinine 134 and inorganic phosphate 89 ml/min. Using bicarbonate dialysate no consistent hypoxemia occurred and all dialyses were well tolerated. The L-3 PC and L-5 PC dialyzers induced more leukopenia and C3a generation than PAN but less than Cuprophan. Thus these new dialyzers offer conventional mass transfer and ultrafiltration rates with much improved membrane biocompatibility.

摘要

在6名慢性透析志愿者患者中,研究了两种新型平行平板透析器(使用面积为0.8平方米的聚碳酸酯(PC)膜(L-3 PC)和1.1平方米的PC膜(L-5 PC))的物理、溶质转运和生物相容性特性。这6名患者中有2名曾因铜仿膜出现严重的首次使用综合征。在严格控制的条件下收集样本。通过收集全部透析液确认了质量传输特性,发现与其他标准透析器相比具有优势。对于L-3 PC,KUF为3.59 ml/h/mmHg TMP - 14.5,标准全血清除率为尿素氮142、肌酐121和无机磷69 ml/min。对于L-5 PC,KUF为6.27 ml/h/mmHg TMP - 28.66,标准清除率为尿素氮155、肌酐134和无机磷89 ml/min。使用碳酸氢盐透析液时,未出现持续的低氧血症,所有透析耐受性良好。L-3 PC和L-5 PC透析器比聚砜膜透析器诱导更多的白细胞减少和C3a生成,但比铜仿膜少。因此,这些新型透析器提供了传统的质量传输和超滤率,同时膜生物相容性有了很大改善。

相似文献

3
Performance of the Lundia PRO dialyzer.伦迪亚PRO透析器的性能
Blood Purif. 1986;4(1-3):40-6. doi: 10.1159/000169425.
7
Clinical investigation of the new dialyzer Gambro PRO-3.
Blood Purif. 1986;4(1-3):47-52. doi: 10.1159/000169426.
9
Biocompatibility of a new polycarbonate dialysis membrane.
Blood Purif. 1986;4(1-3):142-6. doi: 10.1159/000169437.