应用动态碲化镉锌单光子发射计算机断层扫描技术(Tl 和 Tc 示踪剂)联合 N-NH 正电子发射断层扫描评估心脏移植术后心肌血流。

Myocardial Flow Assessment After Heart Transplantation Using Dynamic Cadmium-Zinc-Telluride Single-Photon Emission Computed Tomography With Tl and Tc Tracers and Validated by N-NH Positron Emission Tomography.

机构信息

Department of Nuclear Medicine, National Taiwan University Cancer Center, Taipei (K.-Y.K.).

Graduate Institute of Clinical Medicine, College of Medicine (K.-Y.K.), National Taiwan University, Taipei.

出版信息

Circ Cardiovasc Imaging. 2023 Jun;16(6):e015034. doi: 10.1161/CIRCIMAGING.122.015034. Epub 2023 Jun 14.

Abstract

BACKGROUND

Cardiac allograft vasculopathy (CAV) is an obliterative and diffuse form of vasculopathy and is the most common cause of long-term cardiovascular mortality in heart transplant patients. This study aimed to investigate the diagnostic performance of Tc and Tl tracers in the assessment of CAV using cadmium-zinc-telluride (CZT) single-photon emission computed tomography (SPECT) for myocardial blood flow (MBF) and myocardial flow reserve (MFR) quantification, which was further validated using N-NH positron emission tomography (PET).

METHODS

Thirty-eight patients with prior heart transplantation who underwent CZT SPECT and N-NH PET dynamic scans were included in this study. CZT SPECT with Tc-sestamibi was used in the first 19 patients and Tl-chloride for the remaining patients. To determine the diagnostic accuracy of angiographically defined moderate-to-severe CAV, the analysis included patients who underwent angiographic examinations within 1 year of their second scan.

RESULTS

There were no significant differences in the patient characteristics between the Tl and Tc tracer groups. Both Tl and Tc CZT SPECT-derived stress MBF and MFR values globally and in 3 coronary territories showed good correlations with N-NH PET. The Tl and Tc cohorts did not differ significantly in the correlation coefficients of CZT SPECT versus PET for MBF and MFR, except for stress MBF (Tl:0.95 versus Tc:0.80, =0.03). Tl and Tc CZT SPECT were satisfactory for detecting PET MFR <2.0 (Tl area under the curve, 0.92 [0.71-0.99], Tc area under the curve, 0.87 [0.64-0.97]) and angiographically defined moderate-to-severe CAV, and CZT SPECT results were comparable to that of N-NH PET (CZT area under the curve, 0.90 [0.70-0.99], PET area under the curve, 0.86 [0.64-0.97]).

CONCLUSIONS

This small study suggests that CZT SPECT using Tl and Tc tracers showed comparable MBF and MFR, and the results correlated well with those of N-NH PET. Hence, CZT SPECT with Tl or Tc tracers can be used to detect moderate-to-severe CAV in patients with prior heart transplantation. However, validation using larger studies is warranted.

摘要

背景

心脏同种异体移植血管病(CAV)是一种闭塞性和弥漫性血管病,是心脏移植患者长期心血管死亡率的最常见原因。本研究旨在使用碲锌镉(CZT)单光子发射计算机断层扫描(SPECT)评估 Tc 和 Tl 示踪剂在心肌血流(MBF)和心肌血流储备(MFR)定量评估中的诊断性能,进一步使用 N-NH 正电子发射断层扫描(PET)进行验证。

方法

本研究纳入了 38 例既往接受过心脏移植且接受过 CZT SPECT 和 N-NH PET 动态扫描的患者。在前 19 例患者中使用 Tc- sestamibi 进行 CZT SPECT,其余患者使用 Tl-氯化物。为了确定血管造影定义的中度至重度 CAV 的诊断准确性,分析包括在第二次扫描后 1 年内接受血管造影检查的患者。

结果

Tl 和 Tc 示踪剂组患者的特征无显著差异。Tl 和 Tc CZT SPECT 衍生的应激 MBF 和 MFR 值在全球和 3 个冠状动脉区域均与 N-NH PET 具有良好的相关性。Tl 和 Tc 组之间 CZT SPECT 与 PET 之间的 MBF 和 MFR 相关性系数无显著差异,除了应激 MBF(Tl:0.95 与 Tc:0.80,=0.03)。Tl 和 Tc CZT SPECT 均可满意地检测 PET MFR<2.0(Tl 曲线下面积,0.92[0.71-0.99],Tc 曲线下面积,0.87[0.64-0.97])和血管造影定义的中度至重度 CAV,且 CZT SPECT 结果与 N-NH PET 相当(CZT 曲线下面积,0.90[0.70-0.99],PET 曲线下面积,0.86[0.64-0.97])。

结论

这项小型研究表明,使用 Tl 和 Tc 示踪剂的 CZT SPECT 显示出可比的 MBF 和 MFR,且结果与 N-NH PET 相关性良好。因此,CZT SPECT 可使用 Tl 或 Tc 示踪剂检测既往心脏移植患者的中度至重度 CAV。然而,需要更大规模的研究进行验证。