Chugh Vinay Kumar, Bhatia Navleen Kaur, Shastri Dipti, Shankar Sam Prasanth, Singh Surjit, Sardana Rinkle
All India Institute of Medical Sciences Jodhpur, Department of Dentistry, Rajasthan, India.
King George's Medical University, Department of Orthodontics and Dentofacial Orthopedics, Uttar Pradesh, India.
Turk J Orthod. 2023 Jun 22;36(2):94-100. doi: 10.4274/TurkJOrthod.2022.2022.60.
The aim was to compare the reliability of cephalometric analysis using a smartphone-based application with conventional computer-based imaging software.
Pre-treatment cephalometric radiographs of 50 subjects (26 males, 24 females; mean age, 19.2 years; ±4.2) were traced using the OneCeph® application and Dolphin imaging software®. Two independent observers identified seventeen landmarks and measured fourteen cephalometric measurements at an interval of. Interobserver and intraobserver reliability were evaluated using the intraclass correlation coefficient. Student's t-test was used to compare the means of two measurement methods for observer 1 and observer 2. Additionally, the time taken to complete the cephalometric measurements was also compared between the two methods.
Good (ICC 0.75-0.90) to excellent (ICC 0.90-1.00) interobserver and intraobserver reliability was observed for all hard and soft tissue measurements with both methods. No significant differences were found between the two measurement methods for both observers (p<0.05). OneCeph application took significantly more time to complete the analysis than Dolphin imaging software (p<0.001).
Cephalometric measurements made through a smartphone-based application showed good to excellent interobserver and intraobserver reliability and are comparable with the computer-based software. Therefore, it can be recommended for clinical use. The time taken to complete the cephalometric measurements was more with a smartphone-based application (OneCeph application) compared to computer-based software (Dolphin imaging software).
比较使用基于智能手机的应用程序进行头影测量分析与传统基于计算机的成像软件的可靠性。
使用OneCeph®应用程序和Dolphin成像软件®对50名受试者(26名男性,24名女性;平均年龄19.2岁;±4.2)的治疗前头影测量X线片进行描记。两名独立观察者识别17个标志点,并每隔一段时间测量14个头影测量指标。使用组内相关系数评估观察者间和观察者内的可靠性。采用学生t检验比较观察者1和观察者2两种测量方法的均值。此外,还比较了两种方法完成头影测量所需的时间。
两种方法对所有软硬组织测量的观察者间和观察者内可靠性均良好(ICC 0.75 - 0.90)至优秀(ICC 0.90 - 1.00)。两位观察者的两种测量方法之间均未发现显著差异(p<0.05)。OneCeph应用程序完成分析所需的时间明显比Dolphin成像软件长(p<0.001)。
通过基于智能手机的应用程序进行的头影测量显示出良好至优秀的观察者间和观察者内可靠性,并且与基于计算机的软件相当。因此,可推荐用于临床。与基于计算机的软件(Dolphin成像软件)相比,基于智能手机的应用程序(OneCeph应用程序)完成头影测量所需的时间更长。