Light J K, Pietro T
J Urol. 1986 Sep;136(3):632-5. doi: 10.1016/s0022-5347(17)44999-8.
The neuropathic bladder may exhibit altered function following insertion of the artificial urinary sphincter. Detrusor hyperreflexia worsened in 7 patients following implantation of the device. However, the hyperreflexia may improve spontaneously once healing is complete. Detrusor areflexia with functional diminished compliance may remain unchanged (3 patients) or show an increased graphic gradient (10) postoperatively. These changes may result in hydronephrosis (3 patients) or urinary incontinence depending on the temporal relationship between the cuff pressure and the intravesical filling pressure. Possible explanations for the postoperative deterioration in the compliance are activation of the short neuron system, an increase in the alpha-adrenergic response or sensory receptor adaptation coupled with a severe partial lesion of the long neuron system. Because of the propensity for these changes to occur long-term followup is necessary.
人工尿道括约肌植入后,神经源性膀胱功能可能出现改变。7例患者植入该装置后逼尿肌反射亢进加重。然而,一旦愈合完成,反射亢进可能会自发改善。术后,伴有顺应性功能减退的逼尿肌无反射可能保持不变(3例患者),或显示图形梯度增加(10例)。这些变化可能导致肾积水(3例患者)或尿失禁,具体取决于袖带压力与膀胱内充盈压力之间的时间关系。术后顺应性恶化的可能原因是短神经元系统激活、α-肾上腺素能反应增加或感觉受体适应,同时伴有长神经元系统严重部分损伤。由于这些变化易于长期发生,因此有必要进行长期随访。