The accuracy, timeliness, and reliability of the current methods of heart rate (HR) determination in neonatal resuscitation are debatable, each having its own limitations. We aim to compare three methods of HR assessment: (1) traditional stethoscope, (2) electrocardiogram and traditional stethoscope, and (3) digital stethoscope with loudspeaker amplification of heart sounds.
This was a simulated crossover experiment using a high-fidelity manikin. Each team with a physician, a nurse, and a respiratory therapist performed the resuscitations using the three methods (three different scenarios) in different order. The person controlling the HR via manikin controller was blinded, but the single recorder and the providers were not.
Eighteen resuscitations were completed (six teams of three methods each). The time to first HR recording ( < 0.001), total number of HR recorded ( < 0.001), and time to recognize dips in HR was significantly improved in the digital stethoscope group ( = 0.009).
Use of digital stethoscope with amplification improved documentation of HR and earlier recognition of HR changes.
· Amplified heartbeats during neonatal resuscitation improved documentation.. · Amplified infant heartbeats resulted in earlier recognition of HR changes (increase or decrease).. · Providers using this method had greater satisfaction..
目前在新生儿复苏中确定心率(HR)的准确性、及时性和可靠性存在争议,每种方法都有其自身的局限性。我们旨在比较三种 HR 评估方法:(1)传统听诊器,(2)心电图和传统听诊器,以及(3)具有扬声器放大心音的数字听诊器。
这是一项使用高保真模拟人的交叉实验。每个由一名医生、一名护士和一名呼吸治疗师组成的团队使用三种方法(三种不同的情况)以不同的顺序进行复苏。通过人工心脏控制器控制 HR 的人是盲目的,但单个记录器和提供者不是。
完成了 18 次复苏(每个方法 6 个团队)。首次记录 HR 的时间( < 0.001)、记录的 HR 总数( < 0.001)和识别 HR 下降的时间在数字听诊器组中显著改善( = 0.009)。
使用带放大功能的数字听诊器可改善 HR 记录并更早识别 HR 变化。
·新生儿复苏过程中放大的心跳可改善记录。·放大婴儿的心跳可更早识别 HR 变化(增加或减少)。·使用该方法的提供者满意度更高。