Jargon Jeffrey A, Ginley Ronald A, Sutton Douglas D
National Institute of Standards and Technology, Boulder, CO 80303-3328.
J Res Natl Inst Stand Technol. 1994 Jan-Feb;99(1):19-25. doi: 10.6028/jres.099.003.
An automated linear measurement system (LMS) has been developed to determine the nonlinearity of a tuned 30 MHz power detector over a 6.021 dB range. This detector uses a single thermistor bead design with thermal isolation to obtain nearly linear tracking over a 4:1 change in input power. The nonlinear correction for this change, determined by the LMS, is on the order of 1.00030 (+130 μB) for the detector presently in use. Initial experiments indicate an expanded uncertainty of ±0.138% (±598 μB), which is based upon Type A and Type B components.