Med Biol Eng Comput. 2023 Oct;61(10):2627-2636. doi: 10.1007/s11517-023-02869-0. Epub 2023 Jul 5.
Blood pressure (BP) is the main biomarker for monitoring patients, as its lack of control above values considered normal is a modifiable risk factor for target organ damage. The aim of this study is to evaluate the accuracy of the wearable electronic device photoplethysmography technology (PPG) Samsung Galaxy Watch 4 in determining BP in young patients compared to manual and automatic methods of BP determination. This is a quantitative and cross-sectional study, following validation protocols for wearable devices and BP measurement. It was carried out with twenty healthy young adults, in which BP was measured using four instruments, namely, standard sphygmomanometer device (manual), automatic arm oscillometric device (reference), wrist oscillometric device, and Smartwatch PPG. Eighty systolic blood pressure (SBP) and diastolic blood pressure (DBP) readings were observed. SBP means manual 118 ± 2.20,arm 113 ± 2.54, wrist 118 ± 2.51, and PPG (smartwatch) 113 ± 2.58. Among means, arm and PPG difference is 0.15, arm and wrist 4.95, arm and manual 4.45 wrist with PPG. The mean DBP manual 76.7 ± 1.84, arm 73.6 ± 1.92, wrist 79.3 ± 1.87, and PPG 72.2 ± 1.38. Among means, the difference between the arm and PPG is 1.4 and arm and hand 3.5 mmHg. The correlation shows PPG with manual, arm, and wrist. There was a strong SBP correlation and a moderate DBP correlation between the methods tested, evidencing the accuracy of the PPG smartwatch in relation to the reference method.
血压(BP)是监测患者的主要生物标志物,因为其高于正常值的控制不足是目标器官损伤的可改变风险因素。本研究旨在评估可穿戴电子设备光电容积脉搏波技术(PPG)三星 Galaxy Watch 4 在年轻患者中确定血压的准确性,与手动和自动血压确定方法相比。这是一项定量和横断面研究,遵循可穿戴设备和血压测量的验证方案。在 20 名健康的年轻成年人中进行了这项研究,使用了四种仪器测量血压,即标准血压计设备(手动)、自动臂振荡血压计设备(参考)、手腕振荡血压计和智能手表 PPG。观察了 80 次收缩压(SBP)和舒张压(DBP)读数。SBP 平均值手动为 118 ± 2.20,手臂为 113 ± 2.54,手腕为 118 ± 2.51,PPG(智能手表)为 113 ± 2.58。在平均值中,手臂和 PPG 的差值为 0.15,手臂和手腕的差值为 4.95,手臂和手动的差值为 4.45,手腕和 PPG 的差值为 0.15。DBP 平均值手动为 76.7 ± 1.84,手臂为 73.6 ± 1.92,手腕为 79.3 ± 1.87,PPG 为 72.2 ± 1.38。在平均值中,手臂和 PPG 的差值为 1.4,手臂和手的差值为 3.5 毫米汞柱。相关性显示 PPG 与手动、手臂和手腕相关。所测试方法之间的 SBP 相关性很强,DBP 相关性适中,证明了 PPG 智能手表相对于参考方法的准确性。