Suppr超能文献

家用血压设备制造商之间的袖带尺寸差异:当前患者的困境。

Cuff Size Variation Across Manufacturers of Home Blood Pressure Devices: A Current Patient Dilemma.

机构信息

Division for Heart Disease and Stroke Prevention, Centers for Disease Control and Prevention, Atlanta, Georgia, USA.

出版信息

Am J Hypertens. 2023 Sep 15;36(10):532-535. doi: 10.1093/ajh/hpad060.

Abstract

BACKGROUND

The American Heart Association (AHA) recommends cuff sizes of blood pressure (BP) monitoring devices based on patient arm circumference, which is critical for accurate BP measurement. This study aimed to assess cuff size variation across validated BP devices and to examine the degree of alignment with the AHA recommendations.

METHODS

Data on home BP devices were obtained from the US BP Validated Device Listing website and listed cuff sizes were compared against AHA recommendations: small adult (22-26 cm), adult (27-34 cm), large (35-44 cm), and extra-large (XL) (45-52 cm).

RESULTS

There were 42 home validated BP devices from 13 manufacturers, and none offered cuffs that were aligned with the AHA recommendations. Over half of the devices (22, 52.4%) were compatible with only a broad-range cuff, generally excluding arm sizes larger than 44 cm. Only 5 devices from 4 manufacturers offered a cuff labeled "XL," and of these, only 3 devices had sizes that covered the AHA XL range. Terminology lacked consistency with manufacturers using: different labels to describe the same-sized cuffs (e.g., 22-42 cm was labeled "integrated," "standard," "adult," "large," and "wide range"); the same labels to describe differently sized cuffs (e.g., cuffs labeled "large" were sized 22-42 cm, 32-38 cm, 32-42 cm, 36-45 cm).

CONCLUSIONS

Manufacturers of US home BP devices employ inconsistent terminologies and thresholds for cuff sizes, and sizes were not aligned with AHA recommendations. This lack of standardization could pose challenges for clinicians and patients attempting to select a properly sized cuff to support hypertension diagnosis and management.

摘要

背景

美国心脏协会(AHA)根据患者臂围推荐血压(BP)监测设备的袖带尺寸,这对准确测量血压至关重要。本研究旨在评估经过验证的 BP 设备的袖带尺寸变化,并检查与 AHA 建议的一致性程度。

方法

从美国 BP 验证设备清单网站获取家用 BP 设备的数据,并将列出的袖带尺寸与 AHA 建议进行比较:小成人(22-26cm)、成人(27-34cm)、大(35-44cm)和特大(XL)(45-52cm)。

结果

共有 13 家制造商的 42 种家用经验证的 BP 设备,没有一种设备的袖带与 AHA 建议一致。超过一半的设备(22,52.4%)仅与宽范围袖带兼容,通常排除臂围大于 44cm 的尺寸。只有 4 家制造商的 5 种设备提供标有“XL”的袖带,其中只有 3 种设备的尺寸涵盖了 AHA XL 范围。制造商使用的术语缺乏一致性:用不同的标签来描述相同尺寸的袖带(例如,22-42cm 被标记为“集成”、“标准”、“成人”、“大”和“宽范围”);用相同的标签来描述不同尺寸的袖带(例如,标有“大”的袖带尺寸为 22-42cm、32-38cm、32-42cm、36-45cm)。

结论

美国家用 BP 设备制造商使用的袖带尺寸术语和阈值不一致,并且尺寸与 AHA 建议不一致。这种缺乏标准化可能会给试图选择合适尺寸袖带以支持高血压诊断和管理的临床医生和患者带来挑战。