Suppr超能文献

在腹部肿瘤的多次屏气容积调强弧形治疗期间,将计划的膈肌轮廓叠加在千伏电影图像上的膈肌跟踪系统。

Diaphragm Tracking System That Superimposes Planned Diaphragm Contours on Kilovolt Cine Images During Multiple Breath-Hold Volumetric Modulated Arc Therapy for Abdominal Tumors.

作者信息

Nozawa Yuki, Ohta Takeshi, Katano Atsuto, Yamashita Hideomi, Nakagawa Keiichi

机构信息

Radiology, University of Tokyo Hospital, Tokyo, JPN.

Comprehensive Radiatiion Oncology, University of Tokyo, Tokyo, JPN.

出版信息

Cureus. 2024 Aug 22;16(8):e67540. doi: 10.7759/cureus.67540. eCollection 2024 Aug.

Abstract

We recently published a phantom validation of our diaphragm tracking system, DiaTrak, on an Elekta linear accelerator with an integrated cone-beam computed tomography (CBCT) unit for multiple breath-hold volumetric modulated arc therapy of abdominal tumors, where the diaphragm position was compared between digitally reconstructed radiography (DRR) and kilovolt (kV) projection streaming images by template matching. In the present report, the visual feedback of the diaphragm position was added to the reported system. DICOM-RT diaphragm contour data were additionally exported from a treatment planning system to the DiaTrak PC. Following phantom localization by registering the CBCT to the planning CT images, a projected diaphragm contour was overlaid on each DRR image, whereas another two projected diaphragm contours were superimposed on each kV projection cine image every 180 ms after shifting ±5 mm (set as breath-hold tolerance) in the craniocaudal direction during gantry rotation. It was visually confirmed that the projected diaphragm surface was observed within the two contour lines on the kV cine window. The diaphragm registration errors of the localized phantom were also calculated based on image cross-correlation between the DRR and the projection cine images every 180 ms. It was found that the mean diaphragm registration error was -0.29 mm with a standard deviation of 0.32 mm during the gantry rotation. In conclusion, a new interface for the 5 mm tolerance check was proposed to provide direct visual feedback, thereby giving a sense of assurance to the attending radiotherapy technologists. The calculated diaphragm registration errors were relatively small compared to the tolerance of 5 mm, and therefore it is considered clinically acceptable.

摘要

我们最近发表了关于我们的膈肌跟踪系统DiaTrak的体模验证,该系统用于配备集成锥形束计算机断层扫描(CBCT)单元的医科达直线加速器,用于腹部肿瘤的多次屏气容积调强弧形治疗,通过模板匹配比较了数字重建射线照相(DRR)和千伏(kV)投影流图像之间的膈肌位置。在本报告中,将膈肌位置的视觉反馈添加到了报告的系统中。DICOM-RT膈肌轮廓数据还从治疗计划系统导出到DiaTrak计算机。在通过将CBCT与计划CT图像配准进行体模定位后,将投影的膈肌轮廓叠加在每个DRR图像上,而在机架旋转期间,在头足方向上移动±5 mm(设定为屏气容差)后,每180 ms将另外两个投影的膈肌轮廓叠加在每个kV投影电影图像上。目视确认在kV电影窗口的两条轮廓线内观察到了投影的膈肌表面。还根据每180 ms DRR与投影电影图像之间的图像互相关计算了定位体模的膈肌配准误差。发现在机架旋转期间,平均膈肌配准误差为-0.29 mm,标准差为0.32 mm。总之,提出了一种用于5 mm容差检查的新界面,以提供直接的视觉反馈,从而让放疗技术人员有安全感。与5 mm的容差相比,计算出的膈肌配准误差相对较小,因此在临床上被认为是可接受的。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/aa89/11417412/c0c0419c54dc/cureus-0016-00000067540-i01.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验