Harris W H, McGann W A
J Bone Joint Surg Am. 1986 Sep;68(7):1064-6.
Of the 171 total hip replacements reported on previously that had had a minimum length of follow-up of two years, 117 replacements in 104 patients were analyzed at a minimum of five years postoperatively (average, seventy-four months; range, sixty to ninety-four months) to assess the rate of loosening of the femoral component. At the time of cementing of the femoral component, the medullary canal had been plugged with a bolus of bone cement and then filled with doughy Simplex-P methylmethacrylate in a retrograde fashion using a cement gun. The femoral components, made of a chromium-cobalt alloy, had a rectangular cross-sectional shape to the stem and a medial collar. Three categories of loosening were used: definite (requiring radiographic evidence of migration of the component or the cement), probable (requiring evidence of a complete radiolucent zone at the bone-cement interface on one radiograph or more), and possible (a radiolucent zone at the cement-bone interface of more than 49 per cent but less than 100 per cent on one radiograph or more). One femoral component had been removed for aseptic loosening at another hospital, leaving the patient with a resection arthroplasty. One other (1.7 per cent) was definitely loose. No femoral component was categorized as probably loose, and only two were possibly loose.
在之前报告的171例全髋关节置换术中,有104例患者的117次置换术进行了至少两年的随访,在术后至少五年(平均74个月;范围60至94个月)进行分析,以评估股骨部件的松动率。在股骨部件骨水泥固定时,髓腔先用一团骨水泥堵塞,然后用骨水泥枪以逆行方式填入面团状的Simplex-P甲基丙烯酸甲酯。股骨部件由铬钴合金制成,柄部横截面为矩形,带有内侧颈圈。使用了三类松动情况:明确松动(需要有部件或骨水泥移位的影像学证据)、可能松动(需要在一张或多张X线片上有骨水泥界面处完整的透射线区的证据)和可能存在松动(在一张或多张X线片上骨水泥-骨界面处透射线区超过49%但小于100%)。有一个股骨部件在另一家医院因无菌性松动被取出,患者接受了切除关节成形术。另有一个(1.7%)明确松动。没有股骨部件被归类为可能松动,只有两个可能存在松动。