Suppr超能文献

False alarms in a slit-scan flow system: causes and occurrence rates. Implications and potential solutions.

作者信息

Wheeless L L, Cambier J L, Cambier M A, Kay D B, Wightman L L, Patten S F

出版信息

J Histochem Cytochem. 1979 Jan;27(1):596-9. doi: 10.1177/27.1.374626.

Abstract

A slit-scan technique was developed as a basis for an automated prescreening system for gynecologic cytology. A flow system based on this technique was fabricated and tested and results indicated that false alarms (misclassification of objects or events from normal specimens as abnormal) are the greatest remaining obstacle to development of an automated prescreening instrument. A dual view correlation system was fabricated to provide exact image-contour correlation in flow and permit precise determination of causes and occurrence rates of false alarms. This paper presents data from correlation analyses of 23 normal cytologic specimens. Major causes of false alarms and their implications to automated prescreening are discussed. A technique that would eliminate the majority of false alarms in flow is presented.