使用PetMAP设备评估兔有创和无创血压测量之间的一致性。

Evaluation of agreement between invasive and non-invasive blood pressure measurement using the PetMAP device in rabbits.

作者信息

Sredenšek Jerneja, Žel Jurij, Rocchi Attilio, Gasparik-Küls Nina

机构信息

University Clinic for Companion Animals, Clinical Unit of Anaesthesiology and Perioperative Intensive-Care Medicine, University of Veterinary Medicine Vienna, Vienna, Austria.

Small Animal Clinic, Veterinary Faculty, University of Ljubljana, Ljubljana, Slovenia.

出版信息

Front Vet Sci. 2023 Jul 10;10:1141480. doi: 10.3389/fvets.2023.1141480. eCollection 2023.

Abstract

BACKGROUND

Measurement of arterial blood pressure is recommended in anaesthetized animals to guide perioperative treatment. Invasive blood pressure measurement is considered the gold standard, however it is also technically challenging, requires specialised equipment and carries certain risks. For these reasons, non-invasive blood pressure measurement devices are commonly used and are expected to provide accurate and reliable results. This requirement is particularly true for rabbits, in whom peri-anaesthetic hypotension is commonly observed and in whom perioperative mortality remains disproportionally high. Several authors have compared different non-invasive devices with invasive measurements in rabbits and have reported contrasting results. However, to date no comparison between invasive measurements and the PetMAP device, that has been designed specifically for veterinary medicine, has been reported.

AIM AND HYPOTHESIS

The aim of the study was the comparison of invasive blood pressure measurement with PetMAP in rabbits. We hypothesised that PetMAP would show acceptable agreement with the invasive measurements according to the American College of Veterinary Internal Medicine guidelines.

MATERIALS AND METHODS

Sixteen client-owned rabbits presenting for various surgical interventions were included in the study. Invasive measurements were performed by cannulation of an auricular artery. The PetMAP cuff was applied distal to the elbow according to the manufacturer's guidelines. For each measurement with PetMAP, three invasive blood pressure values were recorded. The mean of the three invasive values was compared with one value measured with PetMAP.

RESULTS

Data collected from 16 rabbits were used for statistical analysis. In the clinical setting, the PetMAP device showed significant overestimation of systolic, diastolic and mean arterial pressure, which were measured in the auricular artery. In addition, the bias was not constant, implying that the device poorly predicted changes in blood pressure.

CONCLUSION

The PetMAP device did not meet any of the American College of Veterinary Internal Medicine recommendations.

摘要

背景

建议在麻醉动物中测量动脉血压以指导围手术期治疗。有创血压测量被认为是金标准,然而它在技术上也具有挑战性,需要专门的设备并且存在一定风险。由于这些原因,无创血压测量设备被广泛使用,并期望能提供准确可靠的结果。对于兔子来说,这一要求尤为重要,因为在兔子中围麻醉期低血压很常见,且围手术期死亡率仍然过高。几位作者已经比较了不同的无创设备与兔子的有创测量结果,并报告了相互矛盾的结果。然而,迄今为止,尚未有关于有创测量与专门为兽医学设计的PetMAP设备之间比较的报道。

目的和假设

本研究的目的是比较兔子有创血压测量与PetMAP测量结果。我们假设根据美国兽医内科学会的指南,PetMAP与有创测量结果将显示出可接受的一致性。

材料和方法

本研究纳入了16只因各种手术干预前来就诊的客户拥有的兔子。通过耳动脉插管进行有创测量。根据制造商的指南,将PetMAP袖带应用于肘部远端。对于每次使用PetMAP的测量,记录三个有创血压值。将三个有创值的平均值与用PetMAP测量的一个值进行比较。

结果

从16只兔子收集的数据用于统计分析。在临床环境中,PetMAP设备显示出对在耳动脉测量的收缩压、舒张压和平均动脉压有明显高估。此外,偏差并不恒定,这意味着该设备对血压变化的预测效果不佳。

结论

PetMAP设备未达到美国兽医内科学会的任何一项建议。