J Clin Med. 2023 Jul 18;12(14):4758. doi: 10.3390/jcm12144758.
Corneal refractive surgery induces high-order aberrations, specifically spherical aberration (SA). These aberrations can have implications when patients later develop cataracts and require the implantation of multifocal intraocular lenses (MIOLs). MIOLs with asymmetric designs pose challenges in predicting outcomes, particularly in these cases. The aim of this study was to evaluate how different values of SA, resulting from various types of refractive surgeries, affect the optical performance of the FEMTIS Comfort intraocular lens.
The through-focus modulation transfer function (TF-MTF) curve and high-contrast images of tumbling E were used as parameters to assess the optical performance of the MIOL. These parameters were measured using an adaptive optics visual simulator.
Increasingly negative values of SA make the MIOL more bifocal, moderating its extended-depth-of-focus characteristic. Conversely, higher positive SA values cause the TF-MTF curve to shift towards positive vergences, leading to worsened distance vision in the +1.00 to +2.00 D range, but improved vision in the +0.50 D to +1.00 D range.
Assessing SA in patients prior to implanting MIOLs with asymmetric designs is necessary for predicting outcomes and making informed decisions based on the visual requirements of patients.
角膜屈光手术会诱发高阶像差,尤其是球差(SA)。当患者日后罹患白内障并需要植入多焦点人工晶状体(MIOL)时,这些像差可能会产生影响。不对称设计的MIOL在预测手术效果时存在挑战,在这些情况下尤为如此。本研究的目的是评估由各种类型屈光手术导致的不同SA值如何影响飞秒舒适型人工晶状体的光学性能。
采用全聚焦调制传递函数(TF-MTF)曲线和翻转E的高对比度图像作为评估MIOL光学性能的参数。这些参数使用自适应光学视觉模拟器进行测量。
SA值越来越负会使MIOL更具双焦特性,减弱其扩展焦深特性。相反,较高的正SA值会导致TF-MTF曲线向正聚散度方向移动,导致在+1.00至+2.00 D范围内的远视力变差,但在+0.50 D至+1.00 D范围内的视力有所改善。
在为不对称设计的MIOL植入患者评估SA对于预测手术效果以及根据患者的视觉需求做出明智决策是必要的。