Suppr超能文献

多焦点和扩展景深人工晶状体在球差条件下的体外光学性能。

In vitro optical performance of multifocal and extended depth-of-focus intraocular lenses in spherical aberration conditions.

机构信息

From the Department of Mechanical Engineering, Pohang University of Science and Technology, Pohang, Gyeongbuk, South Korea (Park, K.H. Kim); Renew Seoul Eye Center, Gangnam-gu, Seoul, South Korea (M.J. Kim).

出版信息

J Cataract Refract Surg. 2022 May 1;48(5):616-622. doi: 10.1097/j.jcrs.0000000000000869.

Abstract

PURPOSE

To evaluate and compare the optical performances of 4 different types of intraocular lenses (IOLs) in various spherical aberration (SA) conditions.

SETTING

POSTECH, Pohang, South Korea.

DESIGN

In vitro laboratory study.

METHODS

A custom optical bench system with adaptive optics was used. A monofocal IOL, a bifocal IOL, a trifocal IOL, and an extended depth-of-focus (EDoF) IOL from Zeiss were evaluated by measuring through-focus modulation transfer function (MTF) as a function of vergence. MTF changes with SA from -0.1 to +0.1 with 0.05 μm step size were analyzed and compared.

RESULTS

In aberration-free conditions, the 4 IOLs showed different MTF curves consistent with their designs. In SA conditions, all the IOLs showed MTF value decreases and the decrease rates at the far focus varied from 28% to 38% per 0.1 μm SAs. The trifocal IOL had low MTF values at the intermediate focus in the noise level with ±0.1 μm SAs.

CONCLUSIONS

All tested IOLs showed MTF decreases with SA in different levels. The trifocal and EDoF IOLs were the most and least sensitive to SA among the evaluated IOLs. The study results might be useful in the selection of IOLs for cataract patients with SA.

摘要

目的

评估和比较 4 种不同类型的人工晶状体(IOL)在不同球差(SA)条件下的光学性能。

地点

韩国浦项科技大学(POSTECH)。

设计

体外实验室研究。

方法

使用具有自适应光学的定制光学台系统。通过测量作为光焦度函数的调制传递函数(MTF)来评估蔡司的单焦点 IOL、双焦点 IOL、三焦点 IOL 和扩展景深(EDoF)IOL。分析并比较了 SA 从-0.1 到+0.1 时,以 0.05μm 步长变化的 MTF 变化。

结果

在无像差条件下,4 种 IOL 显示出与其设计一致的不同 MTF 曲线。在 SA 条件下,所有 IOL 的 MTF 值都下降,每 0.1μm 的 SA 下降率从 28%到 38%不等。在±0.1μm 的 SA 噪声水平下,三焦点 IOL 在中间焦点处的 MTF 值较低。

结论

所有测试的 IOL 在不同水平下均显示出 MTF 随 SA 下降。在评估的 IOL 中,三焦点和 EDoF IOL 对 SA 最敏感,EDoF IOL 最不敏感。该研究结果可能有助于选择有 SA 的白内障患者的 IOL。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验