Barakat Nadim, Browne James A, Cui Quanjun
Department of Orthopaedic Surgery, University of Virginia, Charlottesville, VA, USA.
Arthroplast Today. 2023 Jul 24;22:101161. doi: 10.1016/j.artd.2023.101161. eCollection 2023 Aug.
A 60-year-old man who underwent uncomplicated staged bilateral total hip arthroplasty for femoral head osteonecrosis presented with mechanical catching of his left total hip arthroplasty 3 years after index surgery. Radiographs revealed eccentricity of the left femoral head, concerning the failure of a modern moderately cross-linked polyethylene liner. Catastrophic polyethylene liner failure with significant wear, fragmentation, and femoral head abrasion was noted during revision surgery. The original liner and head were replaced, and the patient has exhibited no complications, pain, or difficulty ambulating 6 months postoperatively. This report highlights one potential novel mechanism for the failure of the Exactech Connexion GXL liner (Exactech Inc., Gainesville, FL), an implant recently reported to have a higher-than-expected failure rate, potentially due to insufficient packaging and increased oxidative processes.