Suppr超能文献

一种现代中度交联聚乙烯髋臼衬垫的灾难性早期失效与破碎

Catastrophic Early Failure and Fragmentation of a Modern Moderately Cross-linked Polyethylene Acetabular Liner.

作者信息

Barakat Nadim, Browne James A, Cui Quanjun

机构信息

Department of Orthopaedic Surgery, University of Virginia, Charlottesville, VA, USA.

出版信息

Arthroplast Today. 2023 Jul 24;22:101161. doi: 10.1016/j.artd.2023.101161. eCollection 2023 Aug.

Abstract

A 60-year-old man who underwent uncomplicated staged bilateral total hip arthroplasty for femoral head osteonecrosis presented with mechanical catching of his left total hip arthroplasty 3 years after index surgery. Radiographs revealed eccentricity of the left femoral head, concerning the failure of a modern moderately cross-linked polyethylene liner. Catastrophic polyethylene liner failure with significant wear, fragmentation, and femoral head abrasion was noted during revision surgery. The original liner and head were replaced, and the patient has exhibited no complications, pain, or difficulty ambulating 6 months postoperatively. This report highlights one potential novel mechanism for the failure of the Exactech Connexion GXL liner (Exactech Inc., Gainesville, FL), an implant recently reported to have a higher-than-expected failure rate, potentially due to insufficient packaging and increased oxidative processes.

摘要

一名60岁男性因股骨头坏死接受了简单的分期双侧全髋关节置换术,在初次手术后3年出现左侧全髋关节置换术的机械性卡顿。X线片显示左侧股骨头偏心,提示现代中度交联聚乙烯衬垫出现故障。翻修手术中发现灾难性的聚乙烯衬垫故障,伴有严重磨损、破碎和股骨头磨损。更换了原来的衬垫和股骨头,患者术后6个月未出现并发症、疼痛或行走困难。本报告强调了Exactech Connexion GXL衬垫(Exactech公司,佛罗里达州盖恩斯维尔)失败的一种潜在新机制,该植入物最近报告的失败率高于预期,可能是由于包装不足和氧化过程增加。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/42ae/10382623/3c2fa23bf173/gr1.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验