BACKGROUND

Historically, there has been excellent survivorship of total hip arthroplasty with newer implants expected to outperform prior generations. Review of our institutional database identified problems with one particular cross-linked polyethylene liner in mid-term follow-up, which has not been reported in the literature. We report on the technical aspects of this liner, as well as the clinical and radiographic factors associated with early failure of this bearing in our practice.

METHODS

A retrospective review of our institutional database was performed from January 2009 to June 2019 of patients who presented with significant osteolysis in the setting of prior total hip arthroplasty with a contemporary polyethylene liner. Demographic and radiographic measures were collected for all patients.

RESULTS

Twelve patients were identified, with nine undergoing revision surgery. All patients demonstrated radiographic osteolysis at the time of revision surgery. The average time to diagnosis of failure was 55.9 months (range: 12-120 months). Average cup position preoperatively was 44 degrees inclination (range: 36-53 degrees) and 21 degrees anteversion (range: 10-34 degrees). Nine patients underwent revision of the acetabular component, with two also undergoing femoral component revision. The Manufacturer and User Facility Device Experience database revealed 22 reported cases of wear-related failure from 2009 to 2019.

CONCLUSIONS

Considering that no identifiable risk factors related to patient demographics or implant position were identified, the Exactech Connexion GXL liner may be prone to a high rate of early failure from wear and severe secondary osteolysis. We recommend close surveillance of patients with this bearing surface.

摘要

背景

从历史上看,新型植入物的全髋关节置换术具有出色的存活率,预计其表现优于前代产品。对我们机构数据库的回顾发现,在中期随访中,有一种特定的交联聚乙烯衬垫存在问题,而这在文献中尚未报道。我们报告了这种衬垫的技术方面,以及我们实践中与这种轴承早期失效相关的临床和影像学因素。

方法

对 2009 年 1 月至 2019 年 6 月期间因先前的全髋关节置换术伴现代聚乙烯衬垫出现明显骨溶解而就诊的患者进行了机构数据库的回顾性研究。收集了所有患者的人口统计学和影像学测量值。

结果

共确定了 12 名患者,其中 9 名接受了翻修手术。所有患者在翻修手术时均显示出影像学骨溶解。平均失败诊断时间为 55.9 个月(范围:12-120 个月)。术前平均杯倾斜度为 44 度(范围:36-53 度),前倾角为 21 度(范围:10-34 度)。9 名患者接受了髋臼部件的翻修,其中 2 名还接受了股骨部件的翻修。制造商和用户设施设备体验数据库显示,2009 年至 2019 年期间,有 22 例与磨损相关的失败报告。

结论

考虑到未确定与患者人口统计学或植入物位置相关的可识别风险因素,Exactech Connexion GXL 衬垫可能容易发生早期磨损和严重继发性骨溶解的高失败率。我们建议密切监测使用这种轴承表面的患者。