Flap perfusion monitoring with an attached surface probe in microvascular reconstruction of the oral cavity.

机构信息

Department of Oral and Maxillofacial Surgery, University Hospital RWTH Aachen, Pauwelsstraße 30, 52074, Aachen, Germany.

出版信息

Clin Oral Investig. 2023 Sep;27(9):5577-5585. doi: 10.1007/s00784-023-05177-x. Epub 2023 Jul 31.

Abstract

OBJECTIVES

Postoperative flap monitoring is essential in oral microvascular reconstruction for timely detection of vascular compromise. This study investigated the use of attached surface probes for the oxygen-2-see (O2C) analysis system (LEA Medizintechnik, Germany) for intraoral flap perfusion monitoring.

MATERIALS AND METHODS

The study included 30 patients who underwent oral reconstruction with a microvascular radial-free forearm flap (RFFF) or anterolateral thigh flap (ALTF) between 2020 and 2022. Flap perfusion was measured with attached (3-mm measurement depth) and unattached surface probes (2- and 8-mm measurement depths) for the O2C analysis system at 0, 12, 24, 36, and 48 h postoperatively. Flap perfusion monitoring with attached surface probes was evaluated for cut-off values for flap blood flow, hemoglobin concentration, and hemoglobin oxygen saturation indicative of vascular compromise and for accuracy and concordance with unattached surface probes.

RESULTS

Three RFFFs were successfully revised, and one ALTF was unsuccessfully revised. The cut-off values indicative of vascular compromise for flap perfusion monitoring with attached surface probes were for RFFF and ALTF: blood flow < 60 arbitrary units (AU) and < 40AU, hemoglobin concentration > 100AU and > 80AU (both > 10% increase), and hemoglobin oxygen saturation < 40% and < 30%. Flap perfusion monitoring with attached surface probes yielded a 97.1% accuracy and a Cohen's kappa of 0.653 (p < 0.001).

CONCLUSIONS

Flap perfusion monitoring with attached surface probes for the O2C analysis system detected vascular compromise accurately and concordantly with unattached surface probes.

CLINICAL RELEVANCE

Attached surface probes for the O2C analysis system are a feasible option for intraoral flap perfusion monitoring.

摘要

目的

口腔微血管重建术后皮瓣监测对于及时发现血管危象至关重要。本研究旨在探讨使用氧-2 见(O2C)分析系统(德国 LEA Medizintechnik)的附着式表面探头进行口腔皮瓣灌注监测。

材料与方法

本研究纳入了 2020 年至 2022 年间接受口腔重建的 30 例患者,使用游离桡侧前臂皮瓣(RFFF)或股前外侧皮瓣(ALTF)。术后 0、12、24、36 和 48 小时,使用附着式(3mm 测量深度)和非附着式表面探头(2mm 和 8mm 测量深度)对 O2C 分析系统进行皮瓣灌注测量。评估附着式表面探头进行皮瓣血流、血红蛋白浓度和血红蛋白氧饱和度监测的截断值,以提示血管危象,并评估其与非附着式表面探头的准确性和一致性。

结果

成功修正了 3 例 RFFF,1 例 ALTF 修正失败。附着式表面探头监测皮瓣灌注的血管危象截断值为:RFFF 和 ALTF 的血流<60 个任意单位(AU)和<40AU,血红蛋白浓度>100AU 和>80AU(均>10%增加),血红蛋白氧饱和度<40%和<30%。附着式表面探头监测皮瓣灌注的准确率为 97.1%,Cohen's kappa 值为 0.653(p<0.001)。

结论

O2C 分析系统的附着式表面探头用于皮瓣灌注监测可准确、一致地检测出血管危象。

临床意义

O2C 分析系统的附着式表面探头是一种可行的口腔皮瓣灌注监测方法。