三种眼压测量方法在眼科疾病患马中的比较。

Comparison of three methods of tonometry in horses presented for ophthalmic disease.

机构信息

New Bolton Center, University of Pennsylvania School of Veterinary Medicine, Kennett Square, Pennsylvania, USA.

出版信息

Vet Ophthalmol. 2023 Sep;26(5):422-427. doi: 10.1111/vop.13135. Epub 2023 Aug 1.

Abstract

OBJECTIVE

To compare the measurement of intraocular pressure in horses with clinical ocular disease using three tonometry devices.

ANIMALS STUDIED

A total of fifty horses were presented to the New Bolton Center Ophthalmology Service.

PROCEDURE

Intraocular pressure was taken on 50 client-owned horses (100 eyes) using the TonoVet, TonoVet Plus and Tono-Pen Avia tonometers. Clinical equine patients included were presented to New Bolton Center for ophthalmic examination. Auriculopalpebral blocks were performed but horses were unsedated.

RESULTS

All three tonometers were found to have strong agreement among them. The strongest agreement was between the TonoVet and the TonoVet Plus with the TonoVet on average 1.330 mmHg less than the TonoVet Plus (standard deviation 4.388 mmHg). This was followed by the TonoVet and the Tono-Pen Avia Vet with the TonoVet on average 2.531 mmHg greater than the Tono-Pen Avia Vet (standard deviation 4.124 mmHg). The weakest agreement was between the TonoVet Plus and the Tono-Pen Avia with the TonoVet Plus on average 3.854 mmHg greater than the Tono-Pen Avia (standard deviation 4.724 mmHg).

CONCLUSIONS

All three tonometers showed strong agreement, however, the TonoVet and the TonoVet Plus carried the strongest agreement with the TonoVet Plus having slightly higher measurements overall compared with the TonoVet. Due to small variations between devices, it is recommended that the same device be used for serial measurements of intraocular pressure. However, all three devices are appropriate to use in horses presented for ophthalmic evaluation.

摘要

目的

比较三种眼压计在患有临床眼病的马中的眼压测量值。

动物研究

共有 50 匹马被送到新博尔顿中心眼科服务处。

程序

使用 TonoVet、TonoVet Plus 和 Tono-Pen Avia 眼压计对 50 匹客户所有的马(100 只眼)进行眼压测量。临床马患者被送到新博尔顿中心进行眼科检查。进行耳-睑阻滞,但马未镇静。

结果

三种眼压计之间均发现具有很强的一致性。TonoVet 与 TonoVet Plus 的一致性最强,TonoVet 的眼压平均比 TonoVet Plus 低 1.330mmHg(标准差为 4.388mmHg)。其次是 TonoVet 与 Tono-Pen Avia Vet,TonoVet 的眼压平均比 Tono-Pen Avia Vet 高 2.531mmHg(标准差为 4.124mmHg)。TonoVet Plus 与 Tono-Pen Avia 的一致性最弱,TonoVet Plus 的眼压平均比 Tono-Pen Avia 高 3.854mmHg(标准差为 4.724mmHg)。

结论

三种眼压计均显示出很强的一致性,但 TonoVet 和 TonoVet Plus 具有最强的一致性,TonoVet Plus 的总体测量值略高于 TonoVet。由于设备之间存在微小差异,建议对眼压进行连续测量时使用同一设备。然而,所有三种设备都适合用于接受眼科评估的马。