Passareli João Victor Goulart Consoni, Nascimento Felipe Franco, Estanho Giovana José Garcia, Ricci Claudia Lizandra, Kanashiro Glaucia Prada, Giuffrida Rogério, Andrade Silvia Franco
Postgraduate Program in Animal Science, UNOESTE, Presidente Prudente, São Paulo, Brazil.
Department of Veterinary Ophthalmology, Veterinary Hospital, UNOESTE, Presidente Prudente, São Paulo, Brazil.
Vet World. 2021 Sep;14(9):2444-2451. doi: 10.14202/vetworld.2021.2444-2451. Epub 2021 Sep 21.
Tonometers are an important instrument for measuring intraocular pressure (IOP) in the diagnosis of glaucoma or uveitis. This study aimed to compare the accuracy of the main types of tonometers with different IOP measurement methodologies in dogs: TonoVet and TonoVet Plus (rebound), Tono-Pen Avia Vet (applanation), and Kowa HA-2 (Goldmann applanation).
IOP was measured in 152 eyes of 76 dogs. A postmortem study was performed by comparing manometry and tonometry values and calculating the correlation coefficient (r), real IOP (manometry) among the tonometers was compared, and an outpatient study was conducted with healthy eyes and eyes with signs of glaucoma and uveitis.
In the postmortem study, the values of r in descending order were Kowa (0.989), TonoVet Plus (0.984), TonoVet (0.981), and Tono-Pen Avia Vet (0.847). The IOP values in mmHg in the study were as follows: Aneroid manometer (16.8±2.5.7), TonoVet (18.1±2.9), TonoVet Plus (20.6±2.3), Tono-Pen Avia Vet (17.1±2.5), and Kowa (16.1±1.7); in outpatient clinics: TonoVet (16.8±3.8), TonoVet Plus (19.2±2.9), Tono-Pen Avia Vet (16.2±2.4), and Kowa (15.0±1.3); glaucoma: TonoVet (30.2±3.5), TonoVet Plus (35.0±6.1), Tono-Pen Avia Vet (29.5±4.2), and Kowa (23.9±5.0); and uveitis: TonoVet (14.2±1.4), TonoVet Plus (17.6±1.9), Tono-Pen Avia Vet (13.7±2.1), and Kowa (12.6±1.7)
There was a strong correlation between IOP values and manometry in all the tonometers. The highest values were obtained with TonoVet Plus and the lowest with Kowa HA-2. All tonometers accurately measured IOP in dogs, including the latest TonoVet Plus, which showed an excellent correlation coefficient.
眼压计是诊断青光眼或葡萄膜炎时测量眼压(IOP)的重要仪器。本研究旨在比较犬中不同眼压测量方法的主要类型眼压计的准确性:TonoVet和TonoVet Plus(回弹式)、Tono-Pen Avia Vet(压平式)以及Kowa HA-2(戈德曼压平式)。
对76只犬的152只眼睛进行眼压测量。通过比较眼压计测量值与压力测量值并计算相关系数(r)进行尸检研究,比较各眼压计之间的实际眼压(压力测量值),并对健康眼睛以及有青光眼和葡萄膜炎体征的眼睛进行门诊研究。
在尸检研究中,r值从高到低依次为Kowa(0.989)、TonoVet Plus(0.984)、TonoVet(0.981)和Tono-Pen Avia Vet(0.847)。研究中的眼压值(mmHg)如下:无液压力计(16.8±2.5.7)、TonoVet(18.1±2.9)、TonoVet Plus(20.6±2.3)、Tono-Pen Avia Vet(17.1±2.5)和Kowa(16.1±1.7);在门诊诊所:TonoVet(16.8±3.8)、TonoVet Plus(19.2±2.9)、Tono-Pen Avia Vet(16.2±2.4)和Kowa(15.0±1.3);青光眼:TonoVet(30.2±3.5)、TonoVet Plus(35.0±6.1)、Tono-Pen Avia Vet(29.5±4.2)和Kowa(23.9±5.0);葡萄膜炎:TonoVet(14.2±1.4)、TonoVet Plus(17.6±1.9)、Tono-Pen Avia Vet(13.7±2.1)和Kowa(12.6±1.7)。
所有眼压计的眼压值与压力测量值之间均存在强相关性。TonoVet Plus获得的值最高,Kowa HA-2获得的值最低。所有眼压计均能准确测量犬的眼压,包括最新的TonoVet Plus,其显示出优异的相关系数。