在评估正常眼压犬的眼前房角时,频域光学相干断层扫描、超声生物显微镜和房角镜的设备间一致性。

Inter-device agreement between spectral domain optical coherence tomography, ultrasound biomicroscopy, and gonioscopy in evaluating the iridocorneal angle in normotensive dogs.

机构信息

Department of Veterinary Clinical Sciences, College of Veterinary Medicine and Research Institute for Veterinary Science, Seoul National University, Seoul 08826, Korea.

出版信息

J Vet Sci. 2023 Jul;24(4):e34. doi: 10.4142/jvs.22241.

Abstract

BACKGROUND

There has not been a thoroughly reported study of the comparison between spectral domain-optical coherence tomography (SD-OCT) with both ultrasound biomicroscopy (UBM) and gonioscopy on the evaluation of the iridocorneal angle (ICA) in dogs.

OBJECTIVES

To investigate the diagnostic value of SD-OCT for the early detection of narrowing ICA by comparing and assessing inter-device agreement in anterior chamber angle (ACA) measurements obtained by SD-OCT and UBM, and ICA evaluations by gonioscopy.

METHODS

A total of 28 eyes from 28 client-owned dogs with normal intraocular pressure were included for examination. The ACA and angle opening distance (AOD) were measured from the SD-OCT and UBM images, and gonioscopy images were analyzed using the ICA grade and ZibWest angle index.

RESULTS

The mean ACA and AOD for SD-OCT were 28.31° ± 5.37° and 658.42 ± 219.90 µm, and for UBM, 28.34° ± 5.82° and 859.29 ± 221.80 µm, respectively. The mean difference in ACA between the average values of SD-OCT and UBM measurements was 0.03° with a 95% limit of agreement (LoA) span of 16.2°, indicating positive agreement; that in AOD was 200.85 µm with a 95% LoA span of 1,110.95 µm, indicating poor agreement. The Pearson correlation coefficient of the ACA of SD-OCT and ZibWest indices of gonioscopy was 0.624, indicating strong agreement; that of UBM and gonioscopy was 0.43, indicating moderate agreement.

CONCLUSIONS

SD-OCT is well tolerated by canine patients due to its non-contact method and might be an alternative option for early screening of ICA narrowing in clinical settings.

摘要

背景

目前尚未有研究全面比较过光谱域光学相干断层扫描(SD-OCT)联合超声生物显微镜(UBM)和房角镜对犬眼前房角(ACA)的评估作用。

目的

通过比较和评估 SD-OCT 和 UBM 测量的前房角(ACA)以及房角镜评估的虹膜角膜角(ICA),探讨 SD-OCT 对早期检测 ICA 变窄的诊断价值。

方法

本研究纳入了 28 只眼压正常的犬的 28 只眼进行检查。从 SD-OCT 和 UBM 图像中测量 ACA 和房角开放距离(AOD),使用 ICA 分级和 ZibWest 角指数分析房角镜图像。

结果

SD-OCT 的平均 ACA 和 AOD 分别为 28.31°±5.37°和 658.42±219.90 µm,UBM 分别为 28.34°±5.82°和 859.29±221.80 µm。SD-OCT 和 UBM 测量的平均值之间的 ACA 平均差异为 0.03°,95%一致性界限(LoA)跨度为 16.2°,表明存在正一致性;AOD 的平均差异为 200.85 µm,95%LoA 跨度为 1,110.95 µm,表明一致性差。SD-OCT 的 ACA 与房角镜的 ZibWest 指数的 Pearson 相关系数为 0.624,表明高度一致;UBM 与房角镜的相关性为 0.43,表明中度一致。

结论

由于 SD-OCT 采用非接触式方法,犬患者易于接受,因此可能是临床早期筛查 ICA 变窄的一种替代方法。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/634b/10404703/9667184ae7a5/jvs-24-e34-g001.jpg