光学相干断层扫描和超声生物显微镜在前房角评估中不同模式的比较。

Comparison of different modes in optical coherence tomography and ultrasound biomicroscopy in anterior chamber angle assessment.

作者信息

Wang Dandan, Pekmezci Melike, Basham Ryan P, He Mingguang, Seider Michael I, Lin Shan C

机构信息

Department of Ophthalmology, University of California, San Francisco, CA, USA.

出版信息

J Glaucoma. 2009 Aug;18(6):472-8. doi: 10.1097/IJG.0b013e31818fb41d.

Abstract

PURPOSE

To compare anterior chamber angle width measurements by low and high-resolution anterior segment optical coherence tomography (AS-OCT) and ultrasound biomicroscopy (UBM).

PATIENTS AND METHODS

Thirty-six eyes of 32 patients with gonioscopically narrow angles from a university-based glaucoma clinic were recruited in the study. All subjects received AS-OCT {in "Anterior Segment Single" mode [low-resolution optical coherence tomography (LOCT)] and "High-resolution Cornea" mode [high-resolution optical coherence tomography (HOCT)]} and UBM examination in 1 or both eyes. Angle opening distance (AOD) and angle recess area of temporal (t) and nasal (n) quadrants were assessed using the customized software on each type of image. Angle measurements from UBM, LOCT, and HOCT images were compared; intraobserver and interobserver measurement reproducibility for each type of image was also evaluated.

RESULTS

Thirty-three eyes of 29 patients were available for analysis. No significant differences were found between LOCT and UBM in the measurement of t-AOD (P=0.350), n-AOD (P=0.106), and n-angle recess area (P=0.360). HOCT measurements were significantly larger than both LOCT and UBM for all parameters (all, P<0.05). HOCT images had better interobserver and intraobserver measurement reproducibility than UBM images.

CONCLUSIONS

LOCT is similar to UBM for most of the studied angle measurements. HOCT tends to give larger measurements than both LOCT and UBM. AS-OCT measurements were more reproducible than those from UBM.

摘要

目的

比较低分辨率和高分辨率眼前节光学相干断层扫描(AS-OCT)及超声生物显微镜检查(UBM)测量前房角宽度的结果。

患者与方法

本研究纳入了某大学青光眼门诊32例房角镜检查显示房角狭窄患者的36只眼。所有受试者均接受了AS-OCT检查(采用“眼前节单模式”[低分辨率光学相干断层扫描(LOCT)]和“高分辨率角膜模式”[高分辨率光学相干断层扫描(HOCT)]),并对1只或双眼进行了UBM检查。使用定制软件在每种图像类型上评估颞侧(t)和鼻侧(n)象限的房角开放距离(AOD)和房角隐窝面积。比较了UBM、LOCT和HOCT图像的房角测量结果;还评估了每种图像类型测量结果在观察者内和观察者间的可重复性。

结果

29例患者的33只眼可供分析。在测量t-AOD(P=0.350)、n-AOD(P=0.106)和n-房角隐窝面积(P=0.360)时,LOCT与UBM之间未发现显著差异。所有参数的HOCT测量值均显著大于LOCT和UBM(均P<0.05)。HOCT图像在观察者间和观察者内测量的可重复性均优于UBM图像。

结论

在大多数所研究的房角测量中,LOCT与UBM相似。HOCT测量值往往大于LOCT和UBM。AS-OCT测量结果比UBM更具可重复性。