Klewin K M, Radius R L
Arch Ophthalmol. 1986 Mar;104(3):395-7. doi: 10.1001/archopht.1986.01050150095035.
Visual field function in the right and left eyes of 31 normal volunteers was evaluated with an automated projection perimeter (OCTOPUS). Serial visual field evaluations were repeated in these same eyes with neutral filters of increasing optical density. We compared the results of threshold determinations with the different neutral filters in place before the examined eye. Significant reduction in threshold sensitivity at several test spots throughout the central 30 degrees visual field was seen with neutral density filters of 0.5 log units or greater. The low level of background illumination of the OCTOPUS perimeter (4.0 apostilbs) may allow relatively minor reduction in light transmission by the ocular media to produce significant changes in the recorded level of threshold sensitivity during visual field evaluation.
使用自动投射视野计(OCTOPUS)对31名正常志愿者的右眼和左眼视野功能进行评估。在这些相同的眼睛中,使用光密度不断增加的中性滤光片重复进行系列视野评估。我们比较了在被检查眼前放置不同中性滤光片时的阈值测定结果。当使用0.5对数单位或更高的中性密度滤光片时,在整个中央30度视野的几个测试点处,阈值敏感性显著降低。OCTOPUS视野计的背景照明水平较低(4.0阿熙提),这可能使得眼内介质对光传输的相对较小减少,就能在视野评估期间使记录的阈值敏感性水平产生显著变化。