Med Sci Monit. 2023 Aug 5;29:e940829. doi: 10.12659/MSM.940829.
BACKGROUND The ankle-brachial index (ABI) is a critical diagnostic test for peripheral artery disease (PAD), albeit requiring technical expertise and dedicated resources. The advent of automatic ABI devices proposes a more accessible approach, necessitating fewer resources and less expertise. This study was conducted to gather data on PAD prevalence and to evaluate the correlation and efficacy of automatic ABI measurements vs traditional Doppler ABI measurements to understand their potential role in primary care settings. MATERIAL AND METHODS ABI measurements were obtained using both the Doppler method and an automatic plethysmographic device (Dopplex ABility, Huntleigh Healthcare). RESULTS Of the 290 participants (mean age 67.6±7.4 years), Doppler ABI method identified 16.8% with abnormal results (<0.9), while the automatic method identified only 5.9%. The mean Doppler ABI was 1.05±0.15, and the mean automatic ABI was 1.12±0.13. The sensitivity of the automatic ABI measurements was 22.2%, and the specificity was 96.8%, with a positive predictive value of 57.1%, and a negative predictive value of 86.9%. Adjustments in the automatic assessment and inclusion of pulse wave velocity enhanced the diagnostic capabilities of the automatic ABI device. CONCLUSIONS While the automatic plethysmographic ABI device may lack the necessary diagnostic capabilities to replace the traditional Doppler ABI device as a standalone test in PAD diagnosis, it could still offer significant value in primary care settings if integrated with adjusted cut-off points and pulse wave velocity analysis.
踝臂指数(ABI)是外周动脉疾病(PAD)的关键诊断测试,尽管需要专业技术和专用资源。自动 ABI 设备的出现提出了一种更易获得的方法,需要更少的资源和更少的专业知识。本研究旨在收集 PAD 患病率的数据,并评估自动 ABI 测量与传统多普勒 ABI 测量的相关性和效果,以了解它们在初级保健环境中的潜在作用。
使用多普勒方法和自动体积描记设备(Dopplex ABility,Huntleigh Healthcare)进行 ABI 测量。
在 290 名参与者中(平均年龄 67.6±7.4 岁),多普勒 ABI 方法发现 16.8%的结果异常(<0.9),而自动方法仅发现 5.9%。多普勒 ABI 的平均值为 1.05±0.15,自动 ABI 的平均值为 1.12±0.13。自动 ABI 测量的灵敏度为 22.2%,特异性为 96.8%,阳性预测值为 57.1%,阴性预测值为 86.9%。自动评估的调整和脉搏波速度的纳入增强了自动 ABI 设备的诊断能力。
虽然自动体积描记 ABI 设备可能缺乏必要的诊断能力,无法替代传统的多普勒 ABI 设备作为 PAD 诊断的独立测试,但如果与调整的截止值和脉搏波速度分析相结合,它仍然可以在初级保健环境中提供重要价值。