Contact Mechanics of Elliptical and Spherical Head Implants during Axial Rotation in Anatomic Total Shoulder Arthroplasty: A Biomechanical Comparison.

作者信息

Muench Lukas N, Slater Maria, Archambault Simon, Berthold Daniel P, Rupp Marco-Christopher, Obopilwe Elifho, Cote Mark P, Mazzocca Augustus D

机构信息

Department of Sports Orthopaedics, Technical University of Munich, 81675 Munich, Germany.

Department of Orthopaedic Surgery, UConn Health, Farmington, CT 06030, USA.

出版信息

J Clin Med. 2023 Jul 26;12(15):4918. doi: 10.3390/jcm12154918.

Abstract

BACKGROUND

Elliptical humeral head implants have been proposed to result in more anatomic kinematics following total shoulder arthroplasty (aTSA). The purpose of this study was to compare glenohumeral contact mechanics during axial rotation using spherical and elliptical humeral head implants in the setting of aTSA.

METHODS

Seven fresh-frozen cadaveric shoulders were utilized for biomechanical testing in neutral (NR), internal (IR), and external (ER) rotation at various levels of abduction (0°, 15°, 30°, 45°, 60°) with lines of pull along each of the rotator cuff muscles. Each specimen underwent the following three conditions: (1) native, and TSA using (2) an elliptical and (3) spherical humeral head implant. Glenohumeral contact mechanics, including contact pressure (CP; kPa), peak contact pressure (PCP; kPa), and contact area (CA; mm), were measured in neutral rotation as well as external and internal rotation using a pressure mapping sensor.

RESULTS

Elliptical head implants showed a significantly lower PCP in ER compared to spherical implants at 0° (Δ-712.0 kPa; = 0.034), 15° (Δ-894.9 kPa; = 0.004), 30° (Δ-897.7 kPa; = 0.004), and 45° (Δ-796.9 kPa; = 0.010) of abduction, while no significant difference was observed in ER at 60° of abduction or at all angles in NR and IR. Both implant designs had similar CA in NR, ER, and IR at all tested angles of abduction ( > 0.05, respectively).

CONCLUSIONS

In the setting of aTSA, elliptical heads showed significantly lower PCP during ER at 0° to 45° of abduction, when compared to spherical head implants. However, in NR and IR, PCP was similar between implant designs. Both designs showed similar CA during NR, ER, and IR at all abduction angles.

LEVEL OF EVIDENCE

basic science; controlled laboratory study.

摘要

背景

椭圆型肱骨头假体已被提出可在全肩关节置换术(aTSA)后产生更符合解剖学的运动学。本研究的目的是比较在aTSA情况下使用球形和椭圆型肱骨头假体进行轴向旋转时的盂肱关节接触力学。

方法

使用7个新鲜冷冻的尸体肩关节进行生物力学测试,在不同外展角度(0°、15°、30°、45°、60°)下进行中立位(NR)、内旋(IR)和外旋(ER),沿每条肩袖肌的拉力线进行。每个标本经历以下三种情况:(1)天然状态,以及使用(2)椭圆型和(3)球形肱骨头假体的TSA。使用压力映射传感器在中立旋转以及外旋和内旋时测量盂肱关节接触力学,包括接触压力(CP;kPa)、峰值接触压力(PCP;kPa)和接触面积(CA;mm)。

结果

在0°(Δ - 712.0 kPa;P = 0.034)、15°(Δ - 894.9 kPa;P = 0.004)、30°(Δ - 897.7 kPa;P = 0.004)和45°(Δ - 796.9 kPa;P = 0.010)外展时,椭圆型头假体在ER中的PCP显著低于球形假体,而在60°外展的ER中或NR和IR的所有角度均未观察到显著差异。在所有测试的外展角度下,两种假体设计在NR、ER和IR中的CA相似(P均> 0.05)。

结论

在aTSA情况下,与球形头假体相比,椭圆型头在0°至45°外展的ER期间显示出显著更低的PCP。然而,在NR和IR中,假体设计之间的PCP相似。两种设计在所有外展角度的NR、ER和IR期间均显示出相似的CA。

证据水平

基础科学;对照实验室研究。