Rotational range of motion of elliptical and spherical heads in shoulder arthroplasty: a dynamic biomechanical evaluation.

机构信息

Department of Orthopaedic Surgery, UConn Health, Farmington, CT, USA.

Department of Orthopaedic Sports Medicine, Technical University of Munich, Munich, Germany.

出版信息

Arch Orthop Trauma Surg. 2022 Jan;142(1):67-76. doi: 10.1007/s00402-020-03587-0. Epub 2020 Aug 31.

Abstract

INTRODUCTION

Elliptical shape humeral head prostheses have been proposed to reflect a more anatomic shoulder replacement. Its effect on the rotational range of motion (ROM) compared to a standard spherical head is still not understood. The purpose was to investigate if there would be a difference in rotational ROM when comparing elliptical and spherical prosthetic heads in a dynamic shoulder model. The authors hypothesized that the use of elliptical heads would result in significantly more rotational ROM compared to the spherical head design.

MATERIALS AND METHODS

Six fresh-frozen, cadaveric shoulders were evaluated using a dynamic shoulder model. After being tested in the native condition, each specimen underwent 6 conditions in the hemiarthroplasty state: (1) matched-fit spherical head, (2) oversized spherical head, (3) undersized spherical head, (4) matched-fit elliptical head, (5) oversized elliptical head, and (6) undersized elliptical head. Following conversion to total shoulder arthroplasty (TSA), the 6 prior conditions were rerun. Each condition was tested at 0°, 30° and 60° of glenohumeral abduction. Rotational ROM was quantified using 3-dimensional tracking, while dynamically applying alternating forces for internal and external rotation via the rotator cuff tendons.

RESULTS

Elliptical and spherical prosthetic heads showed no significant difference in the degree of the total, internal, and external rotational ROM for both the hemiarthroplasty and TSA state. Conversion from hemiarthroplasty to TSA resulted in less degree of total rotational ROM for both head designs in all abduction positions, without reaching statistical significance. There was a significant decrease in total, internal, and external rotational ROM for both elliptical and spherical heads in every replacement condition, when comparing 0° to 30° and 60° of abduction (P < 0.05, respectively).

CONCLUSION

In a dynamic shoulder model, elliptical and spherical prosthetic head designs showed no significant difference in the degree of the total, internal, and external rotational ROM in both the hemiarthroplasty and TSA state.

LEVEL OF EVIDENCE

Controlled laboratory study.

摘要

简介

为了更符合人体解剖结构,人们提出了椭圆形肱骨头假体。但目前仍不清楚其与标准球形头假体相比对旋转活动范围(ROM)的影响。本研究旨在通过动态肩部模型,比较椭圆形和球形假体的旋转 ROM,探讨两者之间是否存在差异。作者假设与球形头设计相比,使用椭圆形头将导致明显更大的旋转 ROM。

材料与方法

本研究使用动态肩部模型评估了 6 个新鲜冷冻的尸体肩部。在原生状态下测试后,每个标本在半肩置换状态下进行 6 种条件测试:(1)匹配球形头;(2)大尺寸球形头;(3)小尺寸球形头;(4)匹配椭圆形头;(5)大尺寸椭圆形头;(6)小尺寸椭圆形头。转换为全肩置换(TSA)后,重新进行了前 6 种条件测试。在盂肱关节外展 0°、30°和 60°时,每种条件均通过 3 维跟踪进行旋转 ROM 定量,同时通过肩袖肌腱动态施加内外旋交替力。

结果

在半肩置换和 TSA 状态下,无论是椭圆形还是球形假体,在总旋转 ROM、内旋和外旋 ROM 方面,两者之间均无显著差异。从半肩置换到 TSA 的转换,两种头型在所有外展位置的总旋转 ROM 度均减少,但无统计学意义。与外展 0°相比,无论是椭圆形还是球形头,在半肩置换和 TSA 状态下,在所有置换条件下,外展 30°和 60°时,总旋转 ROM、内旋和外旋 ROM 度均显著降低(P<0.05)。

结论

在动态肩部模型中,在半肩置换和 TSA 状态下,椭圆形和球形假体头设计在总旋转 ROM、内旋和外旋 ROM 度方面无显著差异。

证据等级

对照实验室研究。