牛血红蛋白基氧载体-201(Hemopure)对四种血液学分析仪的干扰

Interference of bovine hemoglobin-based oxygen carrier-201 (Hemopure) on four hematology analyzers.

作者信息

Bronkhorst-van der Helm Marinke W, Weerkamp Floor, Huisman Albert, Huisman Elise J, Russcher Henk

机构信息

Department of Clinical Chemistry, Erasmus MC, University Medical Center, Rotterdam, The Netherlands.

Department of Clinical Chemistry, Maasstad Hospital, Rotterdam, The Netherlands.

出版信息

Int J Lab Hematol. 2023 Dec;45(6):869-874. doi: 10.1111/ijlh.14146. Epub 2023 Aug 11.

Abstract

INTRODUCTION

Hemoglobin-based oxygen carriers, for example HBOC-201 (Hemopure), are aimed to bridge acute anemia when blood transfusion is not available or refused by the patient. However, since HBOC-201 appears free in plasma, it interferes with laboratory tests. This study presents an overview of HBOC-201 interference on four commonly used hematology analyzers and suggests treatment monitoring possibilities.

METHODS

Blood samples were spiked with therapeutic doses of HBOC-201 and nine hematology parameters were measured with the Sysmex XN-20, Siemens Advia 2120i, Abbott Alinity Hq and Abbot Cell Dyn Sapphire hematology analyzers. The results were compared to control samples and the bias was determined.

RESULTS

Most parameters, including all cell counts, hematocrit and MCV, showed a non-significant bias compared to control. However, the standard, total hemoglobin (Hb) measurement as well as MCH and MCHC showed poor agreement with control, as HBOC-201 was included in this measurement. Yet, the flow cytometry-based Hb method quantified intracellular Hb in spiked samples, excluding HBOC-201.

CONCLUSION

Of all included hematology parameters, only total Hb and the associated MCH and MCHC suffered from interference. In contrast, the flow cytometry-based Hb measurement provided an accurate measure of intracellular Hb. The difference between total Hb and cellular Hb represents the HBOC-201 concentration and can be used to monitor HBOC-201 treatment.

摘要

引言

基于血红蛋白的氧载体,例如HBOC - 201(Hemopure),旨在在无法进行输血或患者拒绝输血时缓解急性贫血。然而,由于HBOC - 201在血浆中呈游离状态,它会干扰实验室检测。本研究概述了HBOC - 201对四种常用血液分析仪的干扰情况,并提出了治疗监测的可能性。

方法

向血样中加入治疗剂量的HBOC - 201,然后使用Sysmex XN - 20、西门子Advia 2120i、雅培Alinity Hq和雅培Cell Dyn Sapphire血液分析仪测量九个血液学参数。将结果与对照样本进行比较并确定偏差。

结果

与对照相比,大多数参数,包括所有细胞计数、血细胞比容和平均红细胞体积,显示出无显著偏差。然而,标准的总血红蛋白(Hb)测量以及平均血红蛋白含量(MCH)和平均血红蛋白浓度(MCHC)与对照的一致性较差,因为该测量中包含了HBOC - 201。然而,基于流式细胞术的Hb方法对加标样本中的细胞内Hb进行了定量,排除了HBOC - 201。

结论

在所有纳入的血液学参数中,只有总Hb以及相关的MCH和MCHC受到干扰。相比之下,基于流式细胞术的Hb测量提供了细胞内Hb的准确测量值。总Hb与细胞内Hb之间的差异代表了HBOC - 201的浓度,可用于监测HBOC - 201治疗。