Right and left ventricular stroke volume measurements with velocity-encoded cine MR imaging: in vitro and in vivo validation.

作者信息

Kondo C, Caputo G R, Semelka R, Foster E, Shimakawa A, Higgins C B

机构信息

Department of Radiology, University of California Medical Center, San Francisco 94143.

出版信息

AJR Am J Roentgenol. 1991 Jul;157(1):9-16. doi: 10.2214/ajr.157.1.2048544.

Abstract

The accuracy of measurements of flow velocity determined by using cine MR phase velocity mapping--velocity-encoded cine (VEC) MR--was assessed by comparing VEC MR data with independent measurements in a flow phantom and in human subjects. Constant flow velocities generated in a phantom (range, 20-408 cm/sec) were determined correctly by VEC MR (r = .997, standard error of the estimate [SEE] = 7.9 cm/sec). Peak systolic velocities in the main pulmonary artery determined by VEC MR correlated well with the measurements obtained by using continuous-wave Doppler echocardiography (r = .91). Stroke volumes measured at the aorta by VEC MR and continuous-wave Doppler imaging also correlated well with each other (r = .80). VEC MR measurements of aortic and pulmonary flow provided left and right ventricular stroke volumes that correlated well with left ventricular stroke volumes determined by short-axis cine MR images (r = .98, SEE = 3.7 ml, and r = .95, SEE = 4.8 ml, respectively). Intra- and interobserver variabilities were small for both left and right ventricular stroke volumes as measured with VEC MR. These results indicate that VEC MR accurately and reproducibly measures aortic and pulmonary flow velocities and volumes in the physiologic range of humans, and can be used to measure right and left ventricular stroke volumes under normal flow conditions.

摘要

通过将电影磁共振相位速度映射(速度编码电影磁共振,VEC MR)所测定的流速测量准确性,与在流动模型及人体受试者中的独立测量结果进行比较来评估。VEC MR能够正确测定模型中产生的恒定流速(范围为20 - 408厘米/秒)(r = 0.997,估计标准误差[SEE] = 7.9厘米/秒)。VEC MR测定的主肺动脉收缩期峰值流速与连续波多普勒超声心动图获得的测量结果相关性良好(r = 0.91)。VEC MR和连续波多普勒成像在主动脉处测量的每搏输出量彼此之间也具有良好的相关性(r = 0.80)。VEC MR对主动脉和肺血流的测量所提供的左右心室每搏输出量,与短轴电影磁共振图像测定的左心室每搏输出量相关性良好(分别为r = 0.98,SEE = 3.7毫升,以及r = 0.95,SEE = 4.8毫升)。用VEC MR测量时,观察者内和观察者间的变异性对于左右心室每搏输出量均较小。这些结果表明,VEC MR能够准确且可重复地测量人体生理范围内的主动脉和肺血流速度及容量,并且可用于在正常血流条件下测量左右心室每搏输出量。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验