评估白内障患者视网膜血氧计图像质量指标。

Evaluation of a retinal oximetry image quality indicator in patients with cataract.

机构信息

Department of Ophthalmology, Rigshospitalet Glostrup, Copenhagen, Denmark.

Department of Clinical Medicine, University of Copenhagen, Copenhagen, Denmark.

出版信息

Acta Ophthalmol. 2024 May;102(3):312-317. doi: 10.1111/aos.15747. Epub 2023 Aug 12.

Abstract

PURPOSE

To evaluate a new automated retinal oximetry image quality indicator with cataract as a clinical model.

METHODS

Sixty-one eyes in 61 patients were imaged by the Oxymap T1 Retinal Oximeter at baseline and 25 eyes were also examined 3 weeks after cataract surgery. Image quality (0-10 on a continuous scale) was compared with standardized AREDS cataract grading and Pentacam lens densitometry. Associations with retinal oximetry measurements and visual acuity were examined.

RESULTS

Image quality correlated with total, nuclear and posterior subcapsular cataract grades (ANOVA, p < 0.05), tended to be associated with lens densitometry and it improved from 4.3 ± 1.4 to 5.7 ± 1.0 (p < 0.05) after cataract surgery. Very low image quality, below 3, led to vessel detection failure in retinal oximetry images. Higher image qualities were linearly associated with higher measured retinal oxygen saturations (r = 0.52 in arteries and r = 0.46 in veins; p < 0.001).

CONCLUSION

Retinal oximetry image quality deteriorated with increasing cataract density and improved after cataract surgery, supporting its use as a measure of optical clarity. The numerical quality indicator demonstrated a threshold below which images of poor optical quality should be discarded. Image quality affects the estimates of retinal oximetry parameters and should therefore be included in future analyses.

摘要

目的

评估一种新的自动化视网膜血氧仪图像质量指标,以白内障作为临床模型。

方法

61 例 61 只眼在基线时用 Oxymap T1 视网膜血氧仪进行成像,其中 25 只眼在白内障手术后 3 周进行了检查。图像质量(0-10 的连续评分)与标准化 AREDS 白内障分级和 Pentacam 晶状体密度计进行比较。并检查与视网膜血氧测量值和视力的相关性。

结果

图像质量与总白内障、核性白内障和后囊下白内障分级相关(ANOVA,p<0.05),与晶状体密度计呈趋势相关,术后从 4.3±1.4 提高到 5.7±1.0(p<0.05)。非常低的图像质量(<3)会导致视网膜血氧仪图像中的血管检测失败。更高的图像质量与更高的测量视网膜氧饱和度呈线性相关(动脉 r=0.52,静脉 r=0.46;p<0.001)。

结论

视网膜血氧仪图像质量随白内障密度的增加而恶化,白内障手术后改善,支持其作为光学清晰度的测量指标。数值质量指标显示了一个阈值,低于该阈值,应丢弃光学质量较差的图像。图像质量会影响视网膜血氧参数的估计,因此应包含在未来的分析中。