Koedam J C, van Dreumel H J, v d Ham W, Steentjes G M, Terlingen J B
Clin Chem. 1986 Oct;32(10):1906-13.
A collaborative study to assign values to two enzyme reference materials (ERMs) was performed by 18 laboratories whose spectrophotometers were checked by us, just before the study. We measured the wavelength accuracy and repeatability, the accuracy and linearity of the absorbance curves, the cuvette pathlength, equilibration time, equilibrium temperature, and a few other variables. Five spectrophotometers exhibited a marked wavelength-dependent nonlinearity. Most instruments were rather slow in bringing the sample to the correct temperature and the final temperature was often too high. In the collaborative study, each participant performed the same manual, well-described methods on four occasions in triplicate, using reagents prepared locally. The relation between the photometric checks and the analytical results is discussed, as well as the treatment of outliers and the effects on the variances. Suggestions are made about various facets of collaborative studies. The values assigned to the two ERMs carry a 95% uncertainty interval of +/- 1-4% of the mean.
18个实验室开展了一项联合研究,旨在为两种酶参考物质(ERM)赋值。在研究开始前,我们对这些实验室的分光光度计进行了检查。我们测量了波长准确性和重复性、吸光度曲线的准确性和线性、比色皿光程、平衡时间、平衡温度以及其他一些变量。五台分光光度计表现出明显的波长依赖性非线性。大多数仪器将样品加热到正确温度的速度相当慢,且最终温度往往过高。在联合研究中,每位参与者使用当地制备的试剂,按照相同的详细说明的手动方法,对四个样本各进行三次重复操作。文中讨论了光度检查与分析结果之间的关系,以及异常值的处理和对方差的影响。针对联合研究的各个方面提出了建议。赋予这两种ERM的值带有95%的不确定区间,为平均值的±1 - 4%。