Dedicated virtual non-contrast images adapted for liver tissue in clinical photon counting CT improve virtual non-contrast imaging in various organs beyond the liver.

机构信息

Department of Radiology, Neuroradiology and Nuclear Medicine, Johannes Wesling University Hospital, Ruhr University Bochum, Bochum, Germany.

出版信息

Eur J Radiol. 2023 Oct;167:111031. doi: 10.1016/j.ejrad.2023.111031. Epub 2023 Aug 9.

Abstract

PURPOSE

Purpose of this study is to re-evaluate the accuracy and diagnostic reliability of virtual non-contrast (VNC) images acquired with the photon-counting computed tomography (PCCT) after an update of the CT scanner software.

METHODS

Fifty-four patients were retrospectively enrolled. VNC images were reconstructed from true non-contrast (TNC) images (VNCn) and contrast-enhanced images in portal venous contrast phase (VNCv). Additionally, a liver-specific VNC (VNCl) was assessed. Quantitative image properties of VNC and TNC images were compared and consistency between VNC images was evaluated. Regions of interest were drawn in the liver, spleen, renal cortex, aorta, muscle and subcutaneous fat.

RESULTS

Attenuation values on all VNC images differed significantly from TNC images in the liver, renal cortex, aorta and fat. A mean offset of <10HU between TNC and all VNC images was found in the liver, spleen and muscle. The comparison of TNC and VNCl images revealed an offset < 10HU in fat. Differences ≤ 10HU between TNC and VNCv and between TNC and VNCl were found in 68%, respectively in 75%. Differences ≤ 15HU were found in 79%, respectively in 92% of all measurements. Differences ≤ 10HU between TNC and VNCn were found in 79% and differences ≤ 15HU in 85%.

CONCLUSION

Although there are statistically significant differences between HU values measured on TNC and VNC images in certain tissues, the minor offsets measured in liver and spleen suggest a good clinical applicability of VNCv and VNCl images. The significantly lower offset in subcutaneous fat on VNCl images suggests a superiority for measurements in adipose tissues.

摘要

目的

本研究旨在评估 CT 扫描仪软件更新后,基于光子计数 CT(PCCT)获得的虚拟非对比(VNC)图像的准确性和诊断可靠性。

方法

回顾性纳入 54 例患者。从真非对比(TNC)图像(VNCn)和门静脉对比期增强图像(VNCv)重建 VNC 图像。另外,评估了肝脏特异性 VNC(VNCl)。比较了 VNC 和 TNC 图像的定量图像特性,并评估了 VNC 图像之间的一致性。在肝脏、脾脏、肾皮质、主动脉、肌肉和皮下脂肪中画出感兴趣区。

结果

所有 VNC 图像的衰减值在肝脏、肾皮质、主动脉和脂肪中与 TNC 图像有显著差异。在肝脏、脾脏和肌肉中,TNC 和所有 VNC 图像之间的平均偏移<10HU。TNC 和 VNCl 图像的比较显示,脂肪中的偏移<10HU。在 TNC 和 VNCv 之间,以及在 TNC 和 VNCl 之间,差异分别在 68%和 75%的情况下<10HU。在 79%和 92%的所有测量中,差异分别在 79%和 92%的情况下<15HU。在 TNC 和 VNCn 之间,差异<10HU 的比例为 79%,差异<15HU 的比例为 85%。

结论

尽管在某些组织中,TNC 和 VNC 图像的 HU 值之间存在统计学上的显著差异,但在肝脏和脾脏中测量的较小偏移表明 VNCv 和 VNCl 图像具有良好的临床适用性。VNCl 图像中皮下脂肪的显著较低偏移表明在脂肪组织测量中具有优越性。