Immersion ultrasound improves the repeatability of shear-wave elastography for measuring median nerve elasticity.

机构信息

Department of Ultrasound, University-Town Hospital of Chongqing Medical University, Chongqing, 401331, China.

Medical Affairs Department, The Fifth People's Hospital of Chongqing, Chongqing, 400030, China.

出版信息

J Orthop Surg Res. 2023 Aug 23;18(1):618. doi: 10.1186/s13018-023-04097-6.

Abstract

OBJECTIVE

To study the factors that influence the measurement of median nerve elasticity, to provide a more reproducible test for the assessment of median nerve elasticity using shear wave elasticity (SWE), and to reduce operator empirical dependence. To compare the repeatability of the median nerve elasticity measurement using immersion SWE with that using contact SWE, analyze the factors affecting SWE measurement, and provide a more repeatable method for doctors without SWE operation experience.

METHODS

Two doctors without SWE operation experience measured the median nerve mean elastic modulus (Emean) at the same position and at different time points on the right wrist of 58 healthy volunteers using immersion and contact ultrasound methods. The intraobserver and interobserver repeatability of measurements was assessed using the interclass correlation coefficient (ICC), while the repeatability was assessed using the Bland-Altman diagram.

RESULTS

The intraobserver and interobserver repeatability of the median nerve elasticity measured via contact SWE by inexperienced operators were classified as good, with ICCs of 0.633 (95% CI 0.380-0.783) and 0.552 (95% CI 0.243-0.735), respectively. The intraobserver and interobserver repeatability of the median nerve elasticity measured by immersion SWE were very good, with ICCs of 0.975 (95% CI 0.958-0.985) and 0.942 (95% CI 0.902-0.966), respectively. The intraobserver and interobserver Bland-Altman diagram of median nerve elasticity measured by immersion SWE showed that 98% of the points fell within the 95% limits of agreement. The intraobserver and interobserver Bland-Altman diagram of median nerve elasticity measured by contact SWE showed that 94% of the points fell within the 95% limits of agreement.

CONCLUSION

Immersion ultrasound can improve the repeatability of median nerve elasticity measurements by inexperienced operators.

摘要

目的

研究影响正中神经弹性测量的因素,为使用剪切波弹性(SWE)评估正中神经弹性提供更具可重复性的测试方法,减少操作者经验依赖性。比较使用浸入式 SWE 和接触式 SWE 测量正中神经弹性的重复性,分析影响 SWE 测量的因素,为没有 SWE 操作经验的医生提供更具可重复性的方法。

方法

两名没有 SWE 操作经验的医生使用浸入式和接触式超声方法,在 58 名健康志愿者右手腕同一位置和不同时间点测量正中神经平均弹性模量(Emean)。采用组内相关系数(ICC)评估测量的观察者内和观察者间重复性,采用 Bland-Altman 图评估重复性。

结果

经验不足的操作者使用接触式 SWE 测量正中神经弹性的观察者内和观察者间重复性均为良好,ICC 分别为 0.633(95%置信区间 0.380-0.783)和 0.552(95%置信区间 0.243-0.735)。使用浸入式 SWE 测量正中神经弹性的观察者内和观察者间重复性均为极好,ICC 分别为 0.975(95%置信区间 0.958-0.985)和 0.942(95%置信区间 0.902-0.966)。浸入式 SWE 测量正中神经弹性的观察者内和观察者间 Bland-Altman 图显示,98%的点落在 95%一致性界限内。接触式 SWE 测量正中神经弹性的观察者内和观察者间 Bland-Altman 图显示,94%的点落在 95%一致性界限内。

结论

浸入式超声可提高经验不足操作者正中神经弹性测量的重复性。