Department of Laboratory Medicine, Chungnam National University College of Medicine, Chungnam National University Hospital, Daejeon, Korea
Cancer Research Institute, Chungnam National University College of Medicine, Daejeon, Korea.
Ann Clin Lab Sci. 2023 Jul;53(4):653-660.
We evaluated the performance of the Alinity hq automated analyzer (Abbott Laboratories, Diagnostics Division, Hematology, Santa Clara, CA, USA). In addition, we determined the reference ranges for the red blood cell (RBC) research parameters.
The precision and stability of the instrument were measured for all complete blood count (CBC) parameters. We compared the CBC results between the Alinity hq and the DxH800 (Beckman Coulter, Miami, FL, USA) and the ADVIA 2120 (Siemens Healthcare Diagnostics, Tarrytown, NY, USA). The white blood cell (WBC) differential results were verified by manual differential counts. We determined the reference ranges of RBC research parameters among healthy adults.
The Alinity hq analyzer demonstrated good within-run and between-day precision for all CBC parameters. The calculated correlation coefficients (r) indicated that Alinity hq-determined values of WBC, RBC, platelet (PLT) counts, hemoglobin (HGB), hematocrit (HCT), and mean corpuscular volume (MCV) were in very good concordance (r>0.95) when compared with results from the DxH800 and the ADVIA 2120. The Alinity hq WBC differential counts were comparable with the manual differential counts, and the results of neutrophil counts by Alinity hq correlated well. Lymphocyte and monocyte count correlated well in samples without blasts.
The Alinity hq presented good analytical performance and showed good correlation compared with other hematology analyzers and manual differential counts.
评估 Alinity hq 自动化分析仪(美国雅培诊断事业部,加利福尼亚州圣克拉拉)的性能。此外,还确定了红细胞(RBC)研究参数的参考范围。
测量了所有全血细胞计数(CBC)参数的仪器精密度和稳定性。比较了 Alinity hq 与 DxH800(贝克曼库尔特,佛罗里达州迈阿密)和 ADVIA 2120(西门子医疗诊断,纽约州塔里敦)的 CBC 结果。通过手动分类计数验证了白细胞(WBC)分类结果。确定了健康成年人 RBC 研究参数的参考范围。
Alinity hq 分析仪在所有 CBC 参数的重复性和日间精密度方面表现良好。计算的相关系数(r)表明,与 DxH800 和 ADVIA 2120 的结果相比,Alinity hq 测定的 WBC、RBC、血小板(PLT)计数、血红蛋白(HGB)、红细胞压积(HCT)和平均红细胞体积(MCV)值具有非常好的一致性(r>0.95)。Alinity hq 的 WBC 分类计数与手动分类计数相当,并且 Alinity hq 中性粒细胞计数的结果相关性良好。在没有原始细胞的样本中,淋巴细胞和单核细胞计数相关性良好。
Alinity hq 具有良好的分析性能,与其他血液学分析仪和手动分类计数相比,相关性良好。