Oral Surg Oral Med Oral Pathol Oral Radiol. 2023 Dec;136(6):769-776. doi: 10.1016/j.oooo.2023.07.006. Epub 2023 Jul 11.
We aimed to map the correlation between thermoluminescent dosimeters (TLDs) and Gafchromic film for measuring absorbed doses and to compare minimum, maximum, and mean absorbed doses over larger regions of interest and at various craniofacial organs and tissues during cone beam computed tomography (CBCT) exposure of the mandibular third molar region.
We positioned TLDs at 75 measurement points in a head phantom. Gafchromic film was cut to the same shape as the 5 levels of the phantom and was placed on top of the TLDs. Both dosimetry methods thus included the surface of each level simultaneously. CBCT scans were made using a 5 × 5 cm field of view and a rotation angle of 200°. Measurements included absorbed dose distributions, doses at all 75 points, and minimum, maximum, and mean doses within organs and tissues.
The correlation of point-dose measurements at all TLD sites with doses measured on film was strong (R = 0.9687), with greatest correlation at lower doses (<2 mGy). Large deviations between TLD and film measurements of minimum and maximum doses and absorbed doses to the organs occurred at all 5 levels. TLD positioning failed to cover several organ sites; for these, only absorbed dose measurements from the film were available.
TLDs were unable to sample dose distributions and gradients accurately. The characteristics of Gafchromic LD-V1 film make it a favorable alternative in dental CBCT dosimetry.
我们旨在绘制热释光剂量计(TLD)与 Gafchromic 胶片之间的相关性,以测量吸收剂量,并比较下颌第三磨牙区域锥形束计算机断层扫描(CBCT)照射期间较大感兴趣区域和各种颅面器官和组织的最小、最大和平均吸收剂量。
我们将 TLD 放置在头模型中的 75 个测量点上。Gafchromic 胶片被切割成与模型的 5 个水平相同的形状,并放置在 TLD 上方。这两种剂量测定方法因此同时包括每个水平的表面。使用 5×5 cm 视野和 200°的旋转角度进行 CBCT 扫描。测量包括吸收剂量分布、所有 75 个点的剂量以及器官和组织内的最小、最大和平均剂量。
所有 TLD 部位的点剂量测量与胶片上测量的剂量之间具有很强的相关性(R=0.9687),在较低剂量(<2 mGy)下相关性最大。在所有 5 个水平,TLD 和胶片测量的最小和最大剂量以及器官吸收剂量之间都存在较大偏差。TLD 定位未能覆盖几个器官部位;对于这些部位,只有胶片的吸收剂量测量值可用。
TLD 无法准确采样剂量分布和梯度。Gafchromic LD-V1 胶片的特性使其成为牙科 CBCT 剂量测定的理想替代品。