Kuswandi Bambang, Irsyad Lukman H, Puspaningtyas Ayik R
Chemo and Biosensors Group, Faculty of Pharmacy, University of Jember, Jl. Kalimantan 37, Jember, East Java, 68121, Indonesia.
Bioimpacts. 2023;13(4):347-353. doi: 10.34172/bi.2023.24195. Epub 2023 Apr 8.
In this work, a flexible, and wearable point-of-care (POC) device integrated on a pain relief patch as wearable colorimetric sensors have been developed for sweat analysis, such as lactic acid, sodium ions, and pH simultaneously. Herein, the patch has still functioned as pain relief, while it allows for sweat monitoring during exercise, and in daily activities.
It was constructed on cotton cloth using wax printing technology (batik stamp) as cloth-based microfluidic devices (CMDs). Here, it uses micro volumes of samples to perform the reaction in the sensing zones, where the sensitive reagents are immobilized so that it can collect and analyze the sweat (lactic acid, sodium ions, and pH) as the model for sweat analytes. The colorimetric analysis was conducted via a smartphone camera by using a free app (Color Grab) for a color image analysis that uses for quantitative analysis or naked eye for semi-qualitative analysis.
The ∆RGB value of the CMDS shows the excellent linear correlation vs analytes concentration, where the coefficient of correlations was found for lactic acid (R = 0.994), sodium ion (R = 0.998), and pH (R = 0.994). The ∆RGB value shows the appropriate color value for the linear correlation of the analyte target concentrations in the sweat samples. Here, the limit of detection (LOD) was found at 45.73 µg/mL for lactic acid and 56.46 µg/mL for sodium ions. The reproducibility was found at 0.79% and 0.89%, for lactic acid and sodium ions respectively.
It was applied for sweat analysis during exercise, and the results show in agreement with the standard methods used in a clinical laboratory.
在本研究中,一种集成在止痛贴片上的灵活可穿戴即时检测(POC)设备作为可穿戴比色传感器被开发出来,用于同时分析汗液中的乳酸、钠离子和pH值。在此,该贴片在发挥止痛作用的同时,还能在运动和日常活动中监测汗液。
它是利用蜡染印花技术(蜡染印章)在棉布上构建的基于布的微流控设备(CMD)。这里,它使用微量样品在传感区域进行反应,传感区域固定有敏感试剂,从而能够收集和分析汗液(乳酸、钠离子和pH值)作为汗液分析物的模型。比色分析通过智能手机摄像头,使用免费应用程序(Color Grab)进行彩色图像分析以进行定量分析,或用肉眼进行半定性分析。
CMD的∆RGB值与分析物浓度呈现出良好的线性相关性,其中乳酸的相关系数为(R = 0.994), 钠离子的相关系数为(R = 0.998),pH值的相关系数为(R = 0.994)。∆RGB值显示了汗液样品中分析物目标浓度线性相关性的合适颜色值。在此,乳酸的检测限为45.73 µg/mL,钠离子的检测限为56.46 µg/mL。乳酸和钠离子的重现性分别为0.79%和0.89%。
该设备应用于运动过程中的汗液分析,结果与临床实验室使用的标准方法一致。