IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, Nîmes, France.
Université de Paris, Assistance Publique Hôpitaux de Paris, Hôpital Européen Georges Pompidou, Department of Radiology, Paris, France.
Med Phys. 2023 Nov;50(11):6828-6835. doi: 10.1002/mp.16701. Epub 2023 Sep 6.
Recently, a second generation of split filter dual-energy CT (SFCT) platform has been developed. The thicknesses of the gold and tin filters used to obtain both low- and high-energy spectra have been changed. These differences in filter thickness may affect the spectral separation between the two spectra and thus the quality of spectral images.
To compare the spectral performance of two Split-Filter Dual-Energy CT systems (SFCT-1 and SFCT-2 ) on virtual monoenergetic images (VMIs) and iodine map.
A Multi-Energy CT phantom was scanned on two SFCT with a tube voltage of 120 kVp for both systems (SFCT-1 -120 and SFCT-2 -120) and 140 kVp only for the second generation (SFCT-2 -140). Acquisitions were performed on the phantom with a CTDI close to 11 mGy. Noise power spectrum (NPS) and task-based transfer function (TTF) were evaluated on VMIs from 40 to 70 keV. A detectability index (d') was computed to assess the detection of two contrast-enhanced lesions on VMIs. Hounsfield Unit (HU) accuracy was assessed on VMIs and the accuracy of iodine concentration was assessed on iodine maps.
For all keV, noise magnitude values were lower with the SFCT-2 -120 than with the SFCT-1 -120 (on average: -22.5 ± 2.9%) and higher with the SFCT-2 -140 than with the SFCT-2 -120 (on average: 25.0 ± 6.2%). Average NPS spatial frequencies (f ) were lower with the SFCT-1 -120 than with the SFCT-2 -120 (-6.0 ± 0.5%) and the SFCT-2 -140 (-3.6 ± 1.6%). Similar TTF values were found for both systems and both kVp for blood and iodine inserts at 2 mg/mL (0.29 ± 0.01 mm ) and at 4 mg/mL (0.31 ± 0.01 mm ). d' values peaked at 40 keV for the SFCT-2 and at 70 keV for the SFCT-1 . Highest d' values were found for the SFCT-2 -120 for both simulated lesions. Accuracy of HU values and iodine concentration was higher with the SFCT-2 than with the SFCT 1 .
Compared to the SFCT-1 , with similar spatial resolution and noise texture values, the SFCT-2 -120 exhibited the lowest values for noise magnitude, the highest detectability index values, and more accurate HU values and iodine concentrations.
最近,第二代分体式滤波双能 CT(SFCT)平台已经开发出来。用于获得低能和高能谱的金和锡滤光片的厚度已经改变。这些滤光片厚度的差异可能会影响两种光谱之间的光谱分离,从而影响光谱图像的质量。
比较两种分体式滤波双能 CT 系统(SFCT-1 和 SFCT-2)在虚拟单能量图像(VMIs)和碘图上的光谱性能。
使用多能量 CT 体模在两种 SFCT 上进行扫描,管电压分别为 120 kVp(SFCT-1-120 和 SFCT-2-120)和 140 kVp(仅第二代 SFCT-2-140)。在接近 11 mGy 的 CTDI 下进行采集。在 40 至 70 keV 的 VMIs 上评估噪声功率谱(NPS)和基于任务的传递函数(TTF)。计算可检测指数(d')以评估在 VMIs 上检测两个对比增强病变的能力。评估 VMIs 上的亨氏单位(HU)准确性和碘浓度的准确性。
在所有 keV 下,SFCT-2-120 的噪声幅度值均低于 SFCT-1-120(平均:-22.5±2.9%),SFCT-2-140 的噪声幅度值均高于 SFCT-2-120(平均:25.0±6.2%)。SFCT-1-120 的平均 NPS 空间频率(f)均低于 SFCT-2-120(-6.0±0.5%)和 SFCT-2-140(-3.6±1.6%)。两种系统和两种管电压下的血液和碘插入物的 TTF 值均相似,在 2 mg/mL 时为 0.29±0.01 mm,在 4 mg/mL 时为 0.31±0.01 mm。SFCT-2 的 d'值在 40 keV 时达到峰值,SFCT-1 的 d'值在 70 keV 时达到峰值。对于两个模拟病变,SFCT-2-120 的 d'值最高。SFCT-2 的 HU 值和碘浓度的准确性均高于 SFCT-1。
与 SFCT-1 相比,SFCT-2-120 具有相似的空间分辨率和噪声纹理值,其噪声幅度值最低,可检测指数值最高,HU 值和碘浓度更准确。