在双源 CT 中可用的四种千伏组合对腹部成像光谱性能的影响:基于任务的体模数据图像质量评估。

Impact of four kVp combinations available in a dual-source CT on the spectral performance of abdominal imaging: A task-based image quality assessment on phantom data.

机构信息

Department of medical imaging, Nîmes Medical Imaging Group, CHU Nimes, Univ Montpellier, Nimes, France.

出版信息

J Appl Clin Med Phys. 2021 Aug;22(8):243-254. doi: 10.1002/acm2.13369. Epub 2021 Jul 26.

Abstract

PURPOSE

To compare the spectral performance of four combinations of kVp available in a third generation dual-source CT (DSCT) on abdominal imaging.

METHODS

An image-quality phantom was scanned with a DSCT using four kVp pairs (tube "A" voltage/tube "B" voltage): 100/Sn150 kVp, 90/Sn150 kVp, 80/Sn150 kVp, and 70/Sn150 kVp, classic parameters and dose level for abdomen examination (CTDI : 11 mGy). The noise power spectrum (NPS) and the task-based transfer function (TTF) of two inserts were computed on virtual monochromatic images (VMIs) at 40/50/60/70 keV and for mixed, low-, and high-kVp images. Detectability index (d') was computed on VMIs and mixed images to model the detection task of liver metastasis (LM) and hepatocellular carcinoma (HCC). Iodine quantification accuracy was assessed using the Root Mean Square Deviation (RMSD ) and the iodine bias (IB).

RESULTS

Noise magnitude decreased by -55%± 0% between 40 and 70 keV for all kVp pairs. Compared to 70/Sn150 kVp, noise magnitude was increased by 9% ± 0% with 80/Sn150 kVp, by 16% ± 1% with 90/Sn150 kVp and by 24%± 1% with 100/Sn150 kVp. The average NPS spatial frequency (f ) shifted toward higher frequencies as energy level increased for all kVp pairs. Lowest f values were found for 70/Sn150 kVp and highest for 100/Sn150 kVp. The value of TTF at 50% (f ) shifted toward lower frequencies with increasing energy level. The highest f  values occurred for 100/Sn150 kVp and the lowest for 80/Sn150 kVp. For both lesions, d' was highest for 70/Sn150 kVp and lowest for 100/Sn150 kVp. Compared to 70/Sn150 kVp, d' decreased by -6% ± 3% with 80/Sn150 kVp, by -11% ± 2% with 90/Sn150 kVp and by -13%± 2% with 100/Sn150 kVp. For all acquisitions, the RSMD and IB were the lowest for 100/Sn150 kVp (0.29 ± 0.10 mg/ml and 0.88 ± 0.30 mg/ml, respectively) and increased when the tube "A" voltage decreased (2.34 ± 0.29 mg/ml for 70/Sn150 kVp and 7.42 ± 0.51 mg/ml respectively).

CONCLUSION

70/Sn150 kVp presented the lowest image noise and highest detectability in VMIs of two small focal liver lesions. 100/Sn150 kVp presented the lowest image noise on mixed images and highest accuracy of iodine quantification in iodine images.

摘要

目的

比较第三代双源 CT(DSCT)中可用的四种 kVp 组合在腹部成像中的光谱性能。

方法

使用 DSCT 对一个图像质量体模进行扫描,使用四种 kVp 对(管 A 电压/管 B 电压):100/Sn150kVp、90/Sn150kVp、80/Sn150kVp 和 70/Sn150kVp,使用经典参数和腹部检查的剂量水平(CTDI:11mGy)。在虚拟单色图像(VMIs)上计算了两个插件的噪声功率谱(NPS)和基于任务的传递函数(TTF),在 40/50/60/70keV 时,以及混合、低和高 kVp 图像时。在 VMIs 和混合图像上计算了检测指标(d'),以模拟肝转移(LM)和肝细胞癌(HCC)的检测任务。使用均方根偏差(RMSD)和碘偏倚(IB)评估碘定量准确性。

结果

所有 kVp 对的噪声幅度在 40keV 至 70keV 之间降低了-55%±0%。与 70/Sn150kVp 相比,80/Sn150kVp 的噪声幅度增加了 9%±0%,90/Sn150kVp 的噪声幅度增加了 16%±1%,100/Sn150kVp 的噪声幅度增加了 24%±1%。随着能量水平的增加,所有 kVp 对的平均 NPS 空间频率(f)向更高频率移动。对于所有 kVp 对,最低 f 值出现在 70/Sn150kVp,最高 f 值出现在 100/Sn150kVp。50%(f)处的 TTF 值随着能量水平的增加而向更低的频率移动。f 值最高的是 100/Sn150kVp,最低的是 80/Sn150kVp。对于两种病变,70/Sn150kVp 的 d'最高,100/Sn150kVp 的 d'最低。与 70/Sn150kVp 相比,80/Sn150kVp 的 d'降低了-6%±3%,90/Sn150kVp 的 d'降低了-11%±2%,100/Sn150kVp 的 d'降低了-13%±2%。对于所有采集,100/Sn150kVp 的 RSMD 和 IB 最低(分别为 0.29±0.10mg/ml 和 0.88±0.30mg/ml),当管 A 电压降低时,RSMD 和 IB 增加(分别为 70/Sn150kVp 的 2.34±0.29mg/ml 和 7.42±0.51mg/ml)。

结论

70/Sn150kVp 在两种小肝病灶的 VMIs 中呈现出最低的图像噪声和最高的可检测性。100/Sn150kVp 在混合图像上呈现出最低的图像噪声,在碘图像上呈现出最高的碘定量准确性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/a213/8364263/28a376d4bf5e/ACM2-22-243-g004.jpg