袖带充气对血压、觉醒、睡眠效率和血氧饱和度下降的影响:VAST初步研究的亚分析

Effect of Cuff Inflation on Blood Pressure, Arousals, Sleep Efficiency, and Desaturations: Sub-Analysis of the VAST Pilot Study.

作者信息

Socrates Thenral, Krisai Philipp, Meienberg Andrea, Mayr Michael, Burkard Thilo, Vischer Annina S

机构信息

Medical Outpatient Department and Hypertension Clinic, ESH Hypertension Centre of Excellence, University Hospital Basel, 4031 Basel, Switzerland.

Department of Cardiology and Cardiovascular Research Institute Basel, University Hospital Basel, University of Basel, 4031 Basel, Switzerland.

出版信息

Diagnostics (Basel). 2023 Sep 7;13(18):2874. doi: 10.3390/diagnostics13182874.

Abstract

The influence of cuff inflations on night-time measurements during 24 h ambulatory blood pressure (BP) measurements is unknown. We investigated the potential effect of cuff inflations on sleep parameters using measurements taken simultaneously with a cuffless device using pulse-transit-time (PTT). On the first day of measurement, standard cuff-based 24 h BP and cuffless measurements were simultaneously performed on the right and left arms (CUFF/PTT-D). In this experiment, 1-2 days after the first measurement, the cuffless device was worn alone (PTT-D). Only data from the cuffless device were analyzed. The following mean sleep parameters were analyzed: mean systolic and diastolic BP, arousals, sleep efficiency, total arousals, arousal per hour, and desaturations. In total, 21 individuals were prospectively enrolled. The mean (SD) age was 47 (±15) years, and 57% were female. The mean systolic asleep BP during CUFF/PTT-D and during PTT-D were 131 (±21) and 131 (±26) mmHg, respectively. The mean diastolic asleep BP values during CUFF/PTT-D and during PTT-D were 80 (±14) and 84 (±14) mmHg, respectively ( = 0.860, = 0.100, respectively). Systolic and diastolic asleep mean difference was 0.1 (±18.0) and -3.6 (±9.8) mmHg, respectively. There were significantly more total arousals during PTT-D ( = 0.042). There were no significant differences seen in sleep efficiency ( = 0.339) or desaturations ( = 0.896) between the two measurement periods. We could not show any significant impact from cuff inflations during sleep, as documented by PTT-D measurements.

摘要

在24小时动态血压测量期间,袖带充气对夜间测量的影响尚不清楚。我们使用基于脉搏传输时间(PTT)的无袖带设备同步测量,研究了袖带充气对睡眠参数的潜在影响。在测量的第一天,在左右手臂上同时进行标准的基于袖带的24小时血压测量和无袖带测量(CUFF/PTT-D)。在该实验中,第一次测量后的1-2天,单独佩戴无袖带设备(PTT-D)。仅分析来自无袖带设备的数据。分析了以下平均睡眠参数:平均收缩压和舒张压、觉醒、睡眠效率、总觉醒次数、每小时觉醒次数和血氧饱和度下降。总共前瞻性招募了21名个体。平均(标准差)年龄为47(±15)岁,57%为女性。在CUFF/PTT-D期间和PTT-D期间的平均收缩期睡眠血压分别为131(±21)和131(±26)mmHg。在CUFF/PTT-D期间和PTT-D期间的平均舒张期睡眠血压值分别为80(±14)和84(±14)mmHg(分别为 = 0.860, = 0.100)。收缩期和舒张期睡眠平均差异分别为0.1(±18.0)和-3.6(±9.8)mmHg。在PTT-D期间总觉醒次数明显更多( = 0.042)。两个测量期间在睡眠效率( = 0.339)或血氧饱和度下降( = 0.896)方面没有显著差异。如PTT-D测量所记录,我们未发现睡眠期间袖带充气有任何显著影响。