Abstract

The effects of the neodymium-yttrium aluminum garnet (Nd-YAG) laser on several commonly used endotracheal tubes were examined. Six different types of tubes were tested for flammability. Three tubes were clear, transparent polyvinylchloride with black lettering on the surface. One was yellow-green silicone rubber with green lettering on the surface. One, which was opaque gray with no printing or marks on it, was designed specifically for use with the carbon dioxide (CO2) laser, and one was opaque red rubber with black letters printed on the surface. The latter was tested with and without a wrapping of reflective aluminum tape. On each tube a clear, unmarked area and an area with printing were tested. In each area the tube was impacted by the laser at three to five points. Pulsed and continuous patterns of laser impact were used. Power ranged from 20 to 50 W. The clear polyvinylchloride tubes were not damaged by a pulsed pattern of exposure in unmarked areas. Segments with black printing or marks were consistently damaged. Continuous laser exposure consistently damaged clear portions of the tube. In clear unmarked areas, silicone rubber tubes were damaged with the pulsed pattern when exposure reached 50 W for 1 s. Areas with green lettering were damaged at the lowest level of exposure. Continuous laser exposure at 30 W damaged clear portions of these tubes after 7 s. Opaque red rubber and opaque gray tubes were damaged at the lowest levels of exposure in both clear and marked areas with pulsed and continuous laser emission. A wrapping of reflective aluminum tape offered no protection.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

研究了钕钇铝石榴石(Nd-YAG)激光对几种常用气管导管的影响。测试了六种不同类型导管的可燃性。三种导管为透明聚氯乙烯材质,表面有黑色字样。一种是黄绿硅胶材质,表面有绿色字样。一种是不透明灰色,无印刷或标记,专为二氧化碳(CO2)激光使用而设计,还有一种是不透明红色橡胶材质,表面印有黑色字母。后者在有和没有反射铝带包裹的情况下进行测试。在每根导管上,对一个无标记的清晰区域和一个有印刷的区域进行测试。在每个区域,导管在三到五个点受到激光照射。使用了脉冲和连续的激光照射模式。功率范围为20至50瓦。透明聚氯乙烯导管在无标记区域的脉冲照射模式下未受损。有黑色印刷或标记的部分持续受损。连续激光照射持续损坏导管的清晰部分。在无标记的清晰区域,当暴露于50瓦持续1秒时,硅胶橡胶导管在脉冲模式下受损。有绿色字样的区域在最低暴露水平下受损。在30瓦连续激光照射7秒后,这些导管的清晰部分受损。不透明红色橡胶导管和不透明灰色导管在有脉冲和连续激光发射的清晰和有标记区域的最低暴露水平下均受损。反射铝带包裹没有提供保护作用。(摘要截选至250字)

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验