This study aimed to investigate the value of multiplexed sensitivity encoding with reversed polarity gradients in improving the quality of diffusion-weighted imaging (DWI) images of the prostate and the diagnostic efficacy of prostate cancer.
Seventy-three patients with prostate disease underwent multiplexed sensitivity encoding with reversed polarity gradients (RPG-MUSE), multiplexed sensitivity encoding (MUSE), and single-shot echo-planar imaging (ssEPI) DWI. Three radiologists performed a qualitative image analysis of the three DWI sequences. Qualitative image analysis included artifact minimization, anatomical detail, and sharpness of prostate edges. Two radiologists measured the signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), geometric distortion rate, and the apparent diffusion coefficient (ADC) values of the prostate disease tissue. Two radiologists jointly performed Prostate Imaging Reporting and Data System scoring of prostate lesions and compared the diagnostic efficacy of the three DWI sequences for prostate cancer.
There was good agreement among radiologists in the evaluation and measurement of the three DWI sequence images (intraclass correlation coefficient >0.75, P < 0.05). The RPG-MUSE DWI images were rated higher than those of MUSE and ssEPI in terms of artifact minimization, anatomical details, and sharpness of prostate edges (P < 0.05). The SNR and CNR of the RPG-MUSE DWI images were higher than those of MUSE and ssEPI (P < 0.05), and the geometric distortion rate was lower than that of the other two sequences (P < 0.05). There were no statistical differences in ADC values between the three DWI sequences (P > 0.05). The diagnostic efficacy of RPG-MUSE and MUSE DWI was higher than that of ssEPI (P < 0.017).
RPG-MUSE can reduce the artifacts and geometric distortion in DWI images of the prostate, improve the SNR and CNR of the images, improve the clarity of anatomical details and boundaries without affecting the measurement of ADC values, has the potential to improve the diagnostic efficacy of prostate lesions, and facilitates the clear display and accurate assessment of prostate lesions.
本研究旨在探讨反转极性梯度多重敏感编码(RPG-MUSE)在提高前列腺弥散加权成像(DWI)图像质量和前列腺癌诊断效能方面的价值。
73 例前列腺疾病患者接受了 RPG-MUSE、MUSE 和单次激发回波平面成像(ssEPI)DWI 检查。3 位放射科医生对 3 种 DWI 序列进行了定性图像分析。定性图像分析包括伪影最小化、解剖细节和前列腺边缘锐利度。2 位放射科医生测量了前列腺病变组织的信噪比(SNR)、对比噪声比(CNR)、几何变形率和表观扩散系数(ADC)值。2 位放射科医生联合对前列腺病变进行前列腺影像报告和数据系统(PI-RADS)评分,并比较了 3 种 DWI 序列对前列腺癌的诊断效能。
3 位放射科医生在 3 种 DWI 序列图像的评估和测量方面具有良好的一致性(组内相关系数>0.75,P<0.05)。在伪影最小化、解剖细节和前列腺边缘锐利度方面,RPG-MUSE DWI 图像的评分均高于 MUSE 和 ssEPI(P<0.05)。RPG-MUSE DWI 图像的 SNR 和 CNR 高于 MUSE 和 ssEPI(P<0.05),几何变形率低于后两者(P<0.05)。3 种 DWI 序列的 ADC 值无统计学差异(P>0.05)。RPG-MUSE 和 MUSE DWI 的诊断效能高于 ssEPI(P<0.017)。
RPG-MUSE 可减少前列腺 DWI 图像的伪影和几何变形,提高图像的 SNR 和 CNR,改善解剖细节和边界的清晰度,同时不影响 ADC 值的测量,有可能提高前列腺病变的诊断效能,有利于前列腺病变的清晰显示和准确评估。