St. Francis Hospital, Memphis, Tennessee, USA.
University of Southern California, Los Angeles, California, USA.
J Appl Clin Med Phys. 2023 Nov;24(11):e14160. doi: 10.1002/acm2.14160. Epub 2023 Oct 4.
The purpose of this guideline is to provide a list of critical performance tests to assist the Qualified Medical Physicist (QMP) in establishing and maintaining a safe and effective quality assurance (QA) program. The performance tests on a linear accelerator (linac) should be selected to fit the clinical patterns of use of the accelerator and care should be given to perform tests which are relevant to detecting errors related to the specific use of the accelerator. Current recommendations for linac QA were reviewed to determine any changes required to those tests highlighted by the original report as well as considering new components of the treatment process that have become common since its publication. Recommendations are made on the acquisition of reference data, routine establishment of machine isocenter, basing performance tests on clinical use of the linac, working with vendors to establish QA tests and performing tests after maintenance and upgrades. The recommended tests proposed in this guideline were chosen based on consensus of the guideline's committee after assessing necessary changes from the previous report. The tests are grouped together by class of test (e.g., dosimetry, mechanical, etc.) and clinical parameter tested. Implementation notes are included for each test so that the QMP can understand the overall goal of each test. This guideline will assist the QMP in developing a comprehensive QA program for linacs in the external beam radiation therapy setting. The committee sought to prioritize tests by their implication on quality and patient safety. The QMP is ultimately responsible for implementing appropriate tests. In the spirit of the report from American Association of Physicists in Medicine Task Group 100, individual institutions are encouraged to analyze the risks involved in their own clinical practice and determine which performance tests are relevant in their own radiotherapy clinics.
本指南的目的是提供一组关键性能测试,以协助合格的医学物理学家(QMP)建立和维护安全有效的质量保证(QA)计划。直线加速器(linac)上的性能测试应根据加速器的临床使用模式进行选择,并应注意进行与特定加速器使用相关的误差检测相关的测试。审查了现行的直线加速器 QA 建议,以确定对原始报告中强调的那些测试进行任何必要的更改,并考虑到自发布以来治疗过程中已成为常见的新组件。就获取参考数据、常规建立机器等中心、根据直线加速器的临床使用情况制定性能测试、与供应商合作建立 QA 测试以及在维护和升级后进行测试提出了建议。本指南中建议的测试是根据指南委员会在评估上次报告中必要的更改后达成的共识选择的。这些测试按测试类别(例如剂量学、机械等)和测试的临床参数进行分组。为每个测试都包含了实施说明,以便 QMP 能够理解每个测试的总体目标。本指南将帮助 QMP 为外照射放射治疗环境中的直线加速器制定全面的 QA 计划。委员会试图根据对质量和患者安全的影响对测试进行优先级排序。QMP 最终负责实施适当的测试。本着美国医学物理学家协会工作组 100 报告的精神,鼓励各个机构分析其临床实践中涉及的风险,并确定在其自己的放射治疗诊所中哪些性能测试是相关的。