Tanaka Osamu, Taniguchi Takuya, Nakaya Shuto, Adachi Kousei, Kiryu Takuji, Makita Chiyoko, Matsuo Masayuki
Department of Radiation Oncology, Asahi University Hospital, Gifu City, Gifu, Japan.
Department of Radiology, Gifu University Hospital, Gifu City, Gifu, Japan.
Rep Pract Oncol Radiother. 2023 Jul 25;28(3):407-415. doi: 10.5603/RPOR.a2023.0040. eCollection 2023.
Stereotactic body radiotherapy (SBRT) is recognized as a curative treatment for oligometastasis. The spinal cord becomes the cauda equina at the lumbar level, and the nerves are located dorsally. Recently, a consensus has been reached that the cauda equina should be contoured as an organ at risk (OAR). Here, we examined the separate contouring benefits for the spinal canal versus the cauda equina only as the OAR.
A medical physicist designed a simulation plan for 10 patients with isolated lumbar metastasis. The OAR was set with three contours: the whole spinal canal, cauda equina only, and cauda equina with bilateral nerve roots. The prescribed dose for the planning target volume (PTV) was 30 Gy/3 fx.
For the constrained QAR doses, D90 and D95 were statistically significant due to the different OAR contouring. The maximum dose () was increased to the spinal canal when the cauda equina max was set to ≤ 20 Gy, but dose hotspots were observed in most cases in the medullary area. The and PTV coverage were negatively correlated for the cauda equina and the spinal canal if was set to ≤ 20 Gy for both.
A portion of the spinal fluid is also included when the spinal canal is set as the OAR. Thus, the PTV coverage rate will be poor if the tumor is in contact with the spinal canal. However, the PTV coverage rate increases if only the cauda equina is set as the OAR.
立体定向体部放疗(SBRT)被认为是寡转移的一种根治性治疗方法。脊髓在腰椎水平变为马尾,神经位于背侧。最近,已达成共识,应将马尾作为危及器官(OAR)进行轮廓勾画。在此,我们研究了仅将脊髓作为OAR与仅将马尾作为OAR进行单独轮廓勾画的益处。
一名医学物理师为10例孤立性腰椎转移患者设计了模拟计划。OAR设置为三种轮廓:整个脊髓、仅马尾以及带有双侧神经根的马尾。计划靶体积(PTV)的处方剂量为30 Gy/3次分割。
对于受限的QAR剂量,由于OAR轮廓勾画不同,D90和D95具有统计学意义。当将马尾最大剂量设置为≤20 Gy时,脊髓的最大剂量()增加,但在大多数情况下,在髓质区域观察到剂量热点。如果两者的均设置为≤20 Gy,则马尾和脊髓的与PTV覆盖率呈负相关。
当将脊髓作为OAR时,也会包括一部分脑脊液。因此,如果肿瘤与脊髓接触,则PTV覆盖率会很差。然而,如果仅将马尾作为OAR,则PTV覆盖率会增加。