Suppr超能文献

Actiwatch Spectrum Plus 在检测儿童户外环境和身体活动中的效用。

Utility of the Actiwatch Spectrum Plus for detecting the outdoor environment and physical activity in children.

机构信息

Department of Optometry, Hadassah Academic College, Jerusalem 9101001, Israel.

Department of Optometry, Hadassah Academic College, Jerusalem 9101001, Israel.

出版信息

J Optom. 2024 Jan-Mar;17(1):100483. doi: 10.1016/j.optom.2023.100483. Epub 2023 Oct 3.

Abstract

PURPOSE

To describe the performance of the Actiwatch Spectrum Plus (Philips, Respironics) for determining real world indoor and outdoor environments and physical activity in children.

METHODS

Children wore the device while performing 10 different activities, ranging from sedentary to vigorous physical-activity, and under different indoor and outdoor conditions. Repeated measures ANOVA was implemented via mixed effects modeling to determine illuminance (lux) and physical activity (counts per 15 s, CP15) across conditions. Receiver operator characteristics (ROC) analysis assessed the accuracy to detect indoor versus outdoor settings.

RESULTS

Illuminance was found to be statistically different across indoor (793 ± 348 lux) and outdoor (4,413 ± 518 lux) conditions (P<.0001), with excellent diagnostic accuracy to detect indoor versus outdoor settings (Area under the ROC Curve, AUC 0.94); 1088 lux was identified as the optimal threshold for outdoor illuminance (sensitivity: 93.0%; specificity: 85.0%). Using published activity ranges, we found that when children were sitting, 94% of the physical-activity readings were classified as sedentary or light. When children were walking, 88% of readings were classified as light, and when children were running, 77% of readings were classified as moderate or vigorous.

CONCLUSION

The Actiwatch Spectrum Plus performed well during real world activities in children, showing excellent diagnostic accuracy at 1088 lux as a threshold to detect indoor versus outdoor environments and in categorizing physical activity.

摘要

目的

描述 Actiwatch Spectrum Plus(飞利浦,瑞斯迈)在儿童现实世界室内和室外环境以及身体活动监测中的性能。

方法

儿童在进行 10 种不同活动时佩戴设备,这些活动从久坐到剧烈运动,以及在不同的室内和室外环境下进行。通过混合效应模型实施重复测量方差分析,以确定不同条件下的光照度(勒克斯)和身体活动(每 15 秒计数,CP15)。接收者操作特征(ROC)分析评估了检测室内与室外环境的准确性。

结果

发现室内(793±348 勒克斯)和室外(4413±518 勒克斯)条件下的光照度存在统计学差异(P<.0001),具有良好的诊断准确性来检测室内与室外环境(ROC 曲线下面积,AUC 为 0.94);1088 勒克斯被确定为室外光照度的最佳阈值(敏感性:93.0%;特异性:85.0%)。使用已发表的活动范围,我们发现当儿童静坐时,94%的身体活动读数被归类为久坐或轻度活动。当儿童行走时,88%的读数被归类为轻度活动,而当儿童跑步时,77%的读数被归类为中度或剧烈活动。

结论

Actiwatch Spectrum Plus 在儿童真实世界活动中表现良好,在 1088 勒克斯作为室内与室外环境的检测阈值以及身体活动的分类方面具有出色的诊断准确性。