Carlsson A, Sharp T, Zetterström T, Ungerstedt U
J Chromatogr. 1986 Oct 24;368(2):299-308. doi: 10.1016/s0021-9673(00)91072-0.
A miniaturized liquid chromatographic system with electrochemical detection (LC-ED) was developed and applied to the analysis of dopamine and its metabolites in dialysate samples collected from the rat brain in vivo. An existing LC-ED system was down-scaled using a 1 mm I.D. small-bore column operated at a reduced flow-rate and with an injection volume of 1 microliter. With the small-bore system the limit of detection for dopamine of ca. 0.06 pg was almost 50 times less than with the conventional system, which represents a two-fold improvement in concentration sensitivity. Miniaturization was accomplished with negligible loss in resolution by using a conventional commercial amperometric detector with minor modifications. The results indicate that a number of useful advantages could be realized by the combination of this small-bore LC-ED system and the in vivo brain dialysis method.
开发了一种带有电化学检测的小型化液相色谱系统(LC-ED),并将其应用于分析从大鼠脑内活体采集的透析液样品中的多巴胺及其代谢产物。使用内径为1毫米的小口径柱,以降低的流速和1微升的进样体积,对现有的LC-ED系统进行了缩小规模。对于小口径系统,多巴胺的检测限约为0.06皮克,几乎比传统系统低50倍,这意味着浓度灵敏度提高了两倍。通过对传统商用安培检测器进行微小修改,在分辨率损失可忽略不计的情况下实现了小型化。结果表明,这种小口径LC-ED系统与活体脑透析方法相结合可实现许多有用的优点。