Evaluation of retinal macular oxygen saturation in healthy controls can aid in understanding the pathological changes seen in similar locations of those with vascular diseases like diabetes.
The aim of this study was to determine the test-retest repeatability of localised retinal oximetry measurements in the macula on the Zilia Oximeter within healthy subjects of different races, 18-40 years old. Oxygen saturation was measured between three time points within the same locations of the right eye.
Twenty seven subjects were included (aged 26.3 ± 3.6 years). All were confirmed to have healthy retinas and at least 6/9 vision. Oximetry measurements were taken using the Zilia to acquire local oxygen saturation measurements (300 µm diameter) at four points 3.1 degrees from the fovea in the superior/temporal, superior/nasal, inferior/temporal, and inferior/nasal locations. Oximetry measurements were taken twice on the same day 20 minutes apart and then again 1-2 weeks later. Oximetry data was analysed with intraclass correlation between visits. To assess intrasubject repeatability, the Bland-Altman repeatability coefficient and coefficient of variation were calculated.
Average Intraclass correlation for the three acquisition times of the right eye was 0.78. The averaged intrasubject repeatability coefficient for the three acquisition times was 8.4. The averaged coefficient of variation was 5.4%.
The Zilia oximeter has good macular test-retest repeatability; however, multiple measurements may be needed to ensure accuracy.
评估健康对照者的视网膜黄斑血氧饱和度有助于理解糖尿病等血管疾病患者相似部位的病理变化。
本研究旨在确定 Zilia 血氧仪在不同种族、18-40 岁健康受试者的黄斑区局部血氧测量的测试-重测重复性。在右眼的相同位置,在三个时间点测量血氧饱和度。
纳入 27 名受试者(年龄 26.3±3.6 岁)。所有受试者均被确认为视网膜健康,且至少有 6/9 的视力。使用 Zilia 血氧仪进行血氧测量,以获取距黄斑 3.1 度的四个位置(上方/颞侧、上方/鼻侧、下方/颞侧和下方/鼻侧)的局部血氧饱和度测量值(300μm 直径)。在同一天两次测量,间隔 20 分钟,然后在 1-2 周后再次测量。使用组内相关系数分析血氧测量值。为了评估个体内重复性,计算了 Bland-Altman 重复性系数和变异系数。
右眼三次采集的平均组内相关系数为 0.78。三次采集的个体内重复性系数平均值为 8.4。平均变异系数为 5.4%。
Zilia 血氧仪具有良好的黄斑测试-重测重复性;然而,为了确保准确性,可能需要多次测量。