Boudebs Georges, Zinoune Julien-Bilal, Cassagne Christophe, Chis Mihaela
Appl Opt. 2023 Oct 10;62(29):7669-7677. doi: 10.1364/AO.500526.
The single-beam -scan thermal lens technique is conducted to evaluate the fluorescence quantum yield of various solutions in the case of high-moderate absorption, considering both scenarios: solutions with substantial fluorescence and solutions with high thermal efficiency but low fluorescence. An analytical calculation is performed to determine the uncertainties associated with the random errors introduced by optical detectors. The results reveal that solutions with low fluorescence lead to a significant error, whereas higher fluorescence can help in decreasing the uncertainty. Additionally, the issue of random errors arising when multiple measurements are needed to accurately estimate the fluorescence of a solution will be discussed in different situations.
采用单光束扫描热透镜技术,在中高吸收情况下评估各种溶液的荧光量子产率,考虑两种情况:具有大量荧光的溶液和具有高热效率但低荧光的溶液。进行了分析计算,以确定与光学探测器引入的随机误差相关的不确定性。结果表明,低荧光溶液会导致显著误差,而较高的荧光有助于降低不确定性。此外,还将讨论在不同情况下,为准确估计溶液荧光而需要进行多次测量时出现的随机误差问题。