Yu Soojoung, Sun Hwa Yeon, Yoo Byungwook
Internaltional Healthcare Center, Soonchunhyang University Seoul Hospital, Seoul, Korea.
Department of Family Medicine, Soonchunhyang University Seoul Hospital, Seoul, Korea.
J Yeungnam Med Sci. 2023 Nov;40(Suppl):S73-S80. doi: 10.12701/jyms.2023.00528. Epub 2023 Oct 20.
The quick and easy nature of point-of-care (POC) testing devices allows regular monitoring of serum lipid levels to increase efficiency. The purpose of this study was to assess a POC lipid analyzer, Barozen Lipid Plus (MICO Biomed Co., Ltd.), which uses capillary blood to measure total cholesterol (TC), triglycerides (TGs), and high-density lipoprotein cholesterol (HDL-C).
Capillary and venous blood samples were collected from 110 participants at a single center in Korea between June 10 and June 26, 2021. TC, TG, and HDL-C measurements using Barozen Lipid Plus were compared with measurements using our reference device, the Roche-Hitachi Cobas 8000 c702 (Hitachi High-Technologies Corporation). This study followed the guidelines of the Clinical and Laboratory Standards Institute and the Clinical Laboratory Improvement Amendments. We surveyed participants regarding the convenience of the POC device using a questionnaire following the completion of blood collection.
When compared to the reference equipment, the measurements obtained using Barozen Lipid Plus were more than 95% satisfactory within TC±10%, TG±25%, and HDL-C±30%. The coefficient of variation in the repeatability testing was within 5% for TC, 5% for TGs, and 7% for HDL-C. The survey results indicated high levels of satisfaction. No adverse events were reported.
These findings suggest that Barozen Lipid Plus is reliable for measuring lipid profiles and can therefore be used to monitor lipid levels at the time and place of patient care.
即时检验(POC)设备快速简便的特性使得能够定期监测血脂水平以提高效率。本研究的目的是评估一种POC血脂分析仪,即Barozen Lipid Plus(MICO Biomed有限公司),该分析仪使用毛细血管血来测量总胆固醇(TC)、甘油三酯(TGs)和高密度脂蛋白胆固醇(HDL-C)。
2021年6月10日至6月26日期间,在韩国的一个单一中心从110名参与者中采集了毛细血管血和静脉血样本。将使用Barozen Lipid Plus测量的TC、TG和HDL-C结果与使用我们的参考设备罗氏日立Cobas 8000 c702(日立高新技术公司)测量的结果进行比较。本研究遵循了临床和实验室标准协会以及临床实验室改进修正案的指导原则。在采血完成后,我们使用问卷对参与者关于POC设备的便利性进行了调查。
与参考设备相比,使用Barozen Lipid Plus获得的测量结果在TC±10%、TG±25%和HDL-C±30%范围内的满意度超过95%。重复性测试中的变异系数对于TC在5%以内,对于TGs在5%以内,对于HDL-C在7%以内。调查结果显示满意度很高。未报告不良事件。
这些发现表明Barozen Lipid Plus在测量血脂谱方面是可靠的,因此可用于在患者护理的时间和地点监测血脂水平。