Linder Pia M, Lan Wenhong, Trautwein Nils F, Brosch-Lenz Julia, von Beschwitz Sebastian, Kupferschläger Jürgen, Reischl Gerald, Grözinger Gerd, Dittmann Helmut, la Fougère Christian, Schmidt Fabian P
Department of Nuclear Medicine and Clinical Molecular Imaging, University Hospital Tuebingen, 72076 Tuebingen, Germany.
Department of Nuclear Medicine, Klinikum Rechts der Isar, Technical University Munich, 81675 Munich, Germany.
Diagnostics (Basel). 2023 Nov 9;13(22):3418. doi: 10.3390/diagnostics13223418.
PET imaging after yttrium-90 (Y-90) radioembolization is challenging because of the low positron fraction of Y-90 (32 × 10). The resulting low number of events can be compensated by the high sensitivity of long axial field-of-view (LAFOV) PET/CT scanners. Nevertheless, the reduced event statistics require optimization of the imaging protocol to achieve high image quality (IQ) and quantification accuracy sufficient for post-treatment dosimetry.
Two phantoms (NEMA IEC and AbdoMan phantoms, mimicking human liver) filled with Y-90 and a 4:1 sphere (tumor)-to-background ratio were scanned for 24 h with the Biograph Vision Quadra (Siemens Healthineers). Eight patients were scanned after Y-90 radioembolization (1.3-4.7 GBq) using the optimized protocol (obtained by phantom studies). The IQ, contrast recovery coefficients (CRCs) and noise were evaluated for their limited and full acceptance angles, different rebinned scan durations, numbers of iterations and post-reconstruction filters. The s-value-based absorbed doses were calculated to assess their suitability for dosimetry.
The phantom studies demonstrate that two iterations, five subsets and a 4 mm Gaussian filter provide a reasonable compromise between a high CRC and low noise. For a 20 min scan duration, an adequate CRC of 56% (vs. 24 h: 62%, 20 mm sphere) was obtained, and the noise was reduced by a factor of 1.4, from 40% to 29%, using the full acceptance angle. The patient scan results were consistent with those from the phantom studies, and the impacts on the absorbed doses were negligible for all of the studied parameter sets, as the maximum percentage difference was -3.89%.
With 2i5s, a 4 mm filter and a scan duration of 20 min, IQ and quantification accuracy that are suitable for post-treatment dosimetry of Y-90 radioembolization can be achieved.
钇-90(Y-90)放射性栓塞后的正电子发射断层显像(PET)具有挑战性,因为Y-90的正电子分数较低(32×10)。长轴视野(LAFOV)PET/CT扫描仪的高灵敏度可弥补由此导致的事件数量减少。然而,事件统计量的减少需要优化成像方案,以实现足以用于治疗后剂量测定的高图像质量(IQ)和定量准确性。
使用Biograph Vision Quadra(西门子医疗)对两个填充Y-90且球体(肿瘤)与背景比例为4:1的体模(NEMA IEC体模和腹部体模)进行24小时扫描。8例患者在接受Y-90放射性栓塞(1.3 - 4.7 GBq)后,采用优化方案(通过体模研究获得)进行扫描。针对有限和全接受角、不同重分组扫描持续时间、迭代次数和重建后滤波器,评估了IQ、对比恢复系数(CRC)和噪声。计算基于s值的吸收剂量,以评估其用于剂量测定的适用性。
体模研究表明,两次迭代、五个子集和4毫米高斯滤波器在高CRC和低噪声之间提供了合理的折衷。对于20分钟的扫描持续时间,使用全接受角时获得了足够的CRC,为56%(与24小时相比:62%,20毫米球体),噪声降低了1.4倍,从40%降至29%。患者扫描结果与体模研究结果一致,对于所有研究的参数集,对吸收剂量的影响可忽略不计,因为最大百分比差异为-3.89%。
采用2i5s、4毫米滤波器和20分钟的扫描持续时间,可实现适用于Y-90放射性栓塞治疗后剂量测定的IQ和定量准确性。