Sensors (Basel). 2023 Nov 10;23(22):9106. doi: 10.3390/s23229106.
Isometric strength testing using a digital dynamometer is reliable for muscle function evaluation. It allows us to objectify manual strength assessment measurement. We tested intra- and inter-observer reliability of a user-friendly efficient digital dynamometer-the Forcemeter-equipped with a computer program to monitor the measurements and to store the data. Abduction, forward flexion, and external and internal rotation of both shoulders were tested three times in 20 healthy volunteers with no record of shoulder trauma. Isometric contracture was recorded in newtons. The first and the third test were carried out by Examiner A (intra-rater reliability); the second test, by Examiner B (inter-rater reliability). Good reliability was shown for intra-class correlation coefficient (ICC) values which mean moderate to high correlations (r = 0.66-0.93) for both examiners. Moderate to high correlations (r = 0.72-0.91) were found for comparisons between the results obtained by Examiner A.
使用数字测力计进行等长力量测试对于肌肉功能评估是可靠的。它使我们能够客观地衡量手动力量评估测量。我们测试了一种用户友好、高效的数字测力计-ForceMeter 的内部和观察者之间的可靠性,该测力计配备了一个计算机程序来监测测量并存储数据。在 20 名没有肩部创伤记录的健康志愿者中,测试了他们的肩部外展、前屈、外旋和内旋三个方向,记录的等长收缩力为牛顿。第一次和第三次测试由考官 A 进行(内部观察者可靠性);第二次测试由考官 B 进行(外部观察者可靠性)。内部观察者的组内相关系数(ICC)值表明具有良好的可靠性,这意味着两者之间具有中度到高度的相关性(r = 0.66-0.93)。考官 A 获得的结果之间的比较也发现了中度到高度的相关性(r = 0.72-0.91)。