健康年轻成年人最大呼吸鼻压测试的可靠性。

Reliability of maximal respiratory nasal pressure tests in healthy young adults.

机构信息

PneumoCardioVascular Lab/HUOL, Hospital Universitário Onofre Lopes, Empresa Brasileira de Serviços Hospitalares (EBSERH) & Departamento de Fisioterapia Universidade Federal do Rio Grande do Norte, Natal, Rio Grande do Norte, Brasil.

Laboratório de Inovação Tecnológica em Reabilitação, Departamento de Fisioterapia, Universidade Federal do Rio Grande do Norte, Natal, Rio Grande do Norte, Brasil.

出版信息

PLoS One. 2023 Nov 29;18(11):e0287188. doi: 10.1371/journal.pone.0287188. eCollection 2023.

Abstract

INTRODUCTION

Sniff nasal inspiratory (SNIP) and expiratory pressure (SNEP) may complement the assessment of respiratory muscle strength. Thus, specifying their reliability is relevant to improving the clinical consistency of both tests.

OBJECTIVE

To assess the reliability of SNIP and SNEP in healthy young adults.

METHODS

This cross-sectional study included self-reported healthy aged 18 to 29 years. SNIP was performed using a plug to occlude one nostril, while SNEP was conducted using a facemask. Participants performed 20 SNIP and SNEP maneuvers with 30-second intervals in between. The intraclass correlation coefficient (ICC), standard error of measurement (SEM), and minimum detectable change (MDC) assessed the reliability of SNIP and SNEP. Analyses were conducted between the highest peak pressure and the first reproducible maneuver in men and women.

RESULTS

The total sample comprised 32 participants: 16 men and 16 women. The ICC, SEM, and MDC for SNIP maneuvers were 0.994 (95%CI 0.988 to 0.997), 1.820 cmH2O, and 5.043 cmH2O, respectively. For SNEP, these parameters were 0.950 (95%CI 0.897 to 0.976), 6.03 cmH2O, and 16.716 cmH2O. The SNIP and SNEP in men showed ICC of 0.992 (95%CI 0.977 to 0.997) and 0.877 (95%CI 0.648 to 0.957), SEM of 2.07 and 7.66 cmH2O, and MDC of 5.74 and 21.23 cmH2O. In women, SNIP and SNEP presented ICC of 0.992 (95%CI 0.977 to 0.997) and 0.957 (95%CI 0.878 to 0.985), SEM of 1.15 and 6.11 cmH2O, and MDC of 3.19 and 16.95 cmH2O. Also, 60% of the highest SNIPs occurred among the 11th and 20th maneuvers in men and women. In men, 55% of the highest SNEPs occurred among the 11th and 20th maneuvers; this value was 50% in women.

CONCLUSION

SNIP and SNEP showed excellent reliability. The reliability of SNIP and SNEP in men was good and excellent, respectively, whereas both tests had excellent reliability in women. Also, women reached the highest peak pressure faster than men in both tests.

摘要

简介

嗅吸口腔吸气(SNIP)和呼气压力(SNEP)可补充呼吸肌力量的评估。因此,明确其可靠性与提高两种测试的临床一致性有关。

目的

评估健康年轻成年人 SNIP 和 SNEP 的可靠性。

方法

本横断面研究纳入了自报年龄在 18 至 29 岁之间的健康成年人。使用塞子阻塞一个鼻孔进行 SNIP,使用面罩进行 SNEP。参与者在 30 秒间隔内进行 20 次 SNIP 和 SNEP 操作。组内相关系数(ICC)、测量误差(SEM)和最小可检测变化(MDC)评估了 SNIP 和 SNEP 的可靠性。在男性和女性中,分别对最高峰压力和第一个可重复的操作进行了分析。

结果

总样本包括 32 名参与者:16 名男性和 16 名女性。SNIP 操作的 ICC、SEM 和 MDC 分别为 0.994(95%CI 0.988 至 0.997)、1.820 cmH2O 和 5.043 cmH2O。对于 SNEP,这些参数分别为 0.950(95%CI 0.897 至 0.976)、6.03 cmH2O 和 16.716 cmH2O。男性的 SNIP 和 SNEP 的 ICC 为 0.992(95%CI 0.977 至 0.997)和 0.877(95%CI 0.648 至 0.957),SEM 为 2.07 和 7.66 cmH2O,MDC 为 5.74 和 21.23 cmH2O。女性的 SNIP 和 SNEP 的 ICC 为 0.992(95%CI 0.977 至 0.997)和 0.957(95%CI 0.878 至 0.985),SEM 为 1.15 和 6.11 cmH2O,MDC 为 3.19 和 16.95 cmH2O。此外,男性和女性中最高 SNIP 的 60%发生在第 11 次和第 20 次操作中。在男性中,最高 SNEP 的 55%发生在第 11 次和第 20 次操作中;在女性中,这一比例为 50%。

结论

SNIP 和 SNEP 显示出极好的可靠性。男性 SNIP 和 SNEP 的可靠性分别为良好和优秀,而两种测试在女性中均具有极好的可靠性。此外,在两种测试中,女性达到最高峰压力的速度均快于男性。